Chapter 2 Flashcards

1
Q

“1.What is the frequency of ACR Accreditation?
(A)Every year
(B)Every 2 years
(C)Every 3 years
(D)Every 5 years

A

“(C) The ACR Accreditation takes place every 3 years. It is recognized as the gold standard and helps assuring the patients that they are getting a high-quality imaging performed and interpreted by qualified individuals. The ACR Accreditation also ensures that the facility meets all equipment quality assurance.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

“2.The material used for the exit port of the mammography tube is necessary because
(A)the intensity of the beam is less on the anode side than on the cathode side
(B)regular glass would harden the emerging beam
(C)the intensity of the beam is more on the anode side than on the cathode side
(D)regular glass would soften the emerging beam”

A

“(B) Mammography uses very low energy x-ray beams, and it is important that the x-ray tube window does not attenuate the low-energy photons, thereby hardening the beam. The proper filter shapes the emission spectrum of the x-ray beam and makes it compatible with the detector and breast characteristics of each patient. In general, mammography units have beryllium as the port window. Borosilicate glass is another option that can be used.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

“3.The intensity of the x-ray beam from the cathode side of the tube is generally higher because
(A)soft characteristic radiation emerges from the anode side
(B)the cathode side is directed to the thickest part of the breast
(C)the heel effect causes variation in the intensity of the x-ray beam
(D)the heel effect increases the intensity of the beam at the anode side”

A

“(C) The heel of the anode reduces the intensity of the x-ray beam only on the anode side. When comparing the beam intensity on the cathode side with the anode side, the intensity on the cathode side will be higher. To compensate for this varying intensity, the mammography tube is turned to position the thickest portion of the breast (posterior) at the cathode end of the tube. In general, the smaller the anode angle, the larger is the heel effect because there is an increased absorption of the rays (Fig. 2–1).”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

“The design of the lip of the compression paddle (both height and angle along the chest wall) affects all of the following except
(A)it prevents the posterior and axillary fat from overlapping the body of the breast
(B)it allows uniform compression of the posterior breast tissue
(C)it helps increasing the structural strength of the compression paddle
(D)it ensures higher compression of the anterior breast tissue

A

“(D) Both the height and angle of the compression paddle make a difference in the final image, and the overall design is considered to increase the structural strength of the compression paddle. A compression device with a rounded or gently sloping posterior edge would not allow uniform compression of the posterior area of the breast. The height of the compression reduces the chance of chest tissue overlapping on the mammogram. The design of the lip has a lesser effect on the anterior aspect of the breast (Fig. 2–2).”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

“The primary goal of compression is to
(A)reduce the object-to-image detector distance (OID) of the lesion
(B)allow uniform penetration of structures within the breast
(C)reduce the possibility of motion during the exposure
(D)reduce the radiation dose to the breast”

A

“B) Compression fulfills all the options (A−D ), but its primary goal is to reduce the breast thickness uniformly, separate breast structures, and allow uniform penetration by the x-ray beam.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

“6.What features of the high-transmission cellular (HTC) grid allow these grids to maintain equal or less radiation dose to the patient compared with linear grids?
(A)These grids use lead as the strips and wood as the interspace material
(B)They have copper as the strips and air as the interspace material
(C)They use aluminum as the strips and air as the interspace material
(D)They use wood as the strips and copper as the interspace material”

A

“(B) HTC grid used in some units has the characteristics of a crossed grid. It can reduce scattered radiation in two directions rather than in one direction as in the linear or focused grid. These grids use copper, rather than lead, as the grid strip, and air, rather than wood or aluminum, as the interspace material. Compared with a linear grid of similar ratio, HTC grids result in equal or less radiation dose to the patient. All grids result in an increased exposure and patient dose but improved contrast (Fig. 2–3).”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

“7.The grid ratio can vary in modern mammography units. A common grid ratio used is
(A)8:1
(B)6:1
(C)5:1
(D)2:1”

A

“(C) Typical grid ratio in mammography is 5:1 (can be as low as 3:1) with frequencies of 30 to 50 line pairs per centimeter (30 lp/cm is most common). Grids in mammography are designed to improve contrast by 15% to 70% depending on the composition and thickness of the breast. Most mammography grids are focused grids with the grid strips angled to match the emerging beam. They are linear grids with carbon fiber or other low-attenuation material as the interspace material. Scatter absorption depends on the grid ratio (grid ratio = height of the lead strips/the distance between the strips [h/d]) and the frequency (number of grid lines per centimeter). Higher ratio means higher scatter absorption but requires higher technical factors. On average, mammography grids have a lower grid ratio compared with general radiography grids (Fig. 2–4).”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

“The chest wall edge of the compression paddle should be aligned just beyond the chest wall edge of the detector to
(A)avoid pushing the patient’s chest away and losing breast tissue
(B)properly position and compress the breast
(C)permit uniform exposure and reduce patient discomfort
(D)avoid projecting the chest wall edge of the paddle on the mammogram

A

“(D) The compression paddle/plate is specifically designed for proper positioning and compression of the breast while reducing discomfort to the patient. The placement of the lip, just beyond the chest wall edge of the detector, prevents the projection of an image of the chest wall edge of the paddle on the mammogram.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

“Which of the following affects focal spot size?
(A)Angle of the anode
(B)A decrease in the source-to-image detector distance (SID)
(C)Decreasing the size of the collimated beam
(D)Changing the relationship between the object-to-image detector distance (OID) and the SID”

A

“(A) The focal spot is the area where electrons strike on the target. In the design based on the line-focus principle, the target is angled allowing a larger area for the electrons to strike while maintaining a small, effective focal spot size. The effective focal spot is the area projected onto the patient or detector. It is also the value quoted when identifying a small or large focal spot size. The smaller the target angle, the smaller the focal spot size. Although the spatial resolution (the ability of the system to visualize individual microcalcifications as separate entities) is related to the focal spot size, changes in the SID, OID, and size of the collimated field do not affect the focal spot size (Fig. 2–5).”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

“Which of the following is the commonly used focal spot size for routine work in mammography?
(A)3 mm
(B)0.3 mm
(C)1 mm
(D)0.1 mm”

A

“(B) Mammography units generally have two focal spot sizes. The large focal spot size may be 0.4 mm or less (generally 0.3 mm), and the small focal spot size ranges from 0.15 to 0.1 mm (generally 0.1 mm). Routine work uses the large focal spot size. The small focal spot size is used for magnification.”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

“11.Which of the following mammography quality control tests is performed as needed?
(A)Phantom image quality
(B)Visual checklist
(C)Repeat analysis
(D)Mammography equipment evaluation (MEE)”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

“12.The general criteria to pass the new American College of Radiology—Digital Mammography (ACR DM) phantom require a minimum of _______ masses.
(A)2
(B)3
(C)4
(D)5”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

“What is the compression thickness indicator test?
(A)A check of the minimum and maximum compression force on automatic compression
(B)A check of the minimum and maximum compression force on manual compression
(C)A check to ensure that the indicated compression thickness is within normal range
(D)A check of the compression on the breast”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

“What is the name given to the test to determine whether the unit has consistent detector signal-to-noise level for all breast thickness?
(A)kVp accuracy
(B)Visual checklist
(C)Acquisition workstation (AWS) QC
(D)AEC test”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

“15.A technologist test that is performed semiannually is
(A)phantom image test
(B)visual checklist
(C)compression force test
(D)repeat analysis”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

“Repeats are
(A)images taken during a breast stereolocalization
(B)images used for QC
(C)images that involve unnecessary exposure to the patient
(D)all discarded images”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

“17.To get a meaningful analysis, the repeat analysis needs a patient volume of at least ____ patients.
(A)200
(B)150
(C)300
(D)250”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

“18.Mammography facilities can receive certification from
1.ACR
2.FDA
3.States as Certifiers (SACs)
(A)1 or 2 only
(B)2 or 3 only
(C)1 or 3 only
(D)1, 2, or 3”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

“19.Aluminum filtration is likely to be matched with __________ targets.
(A)molybdenum
(B)rhodium
(C)tungsten
(D)beryllium”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

“Which of the following visual checklist tests must be performed before clinical use?
(A)Paddles/face shield must not be cracked
(B)Indicators must be working correctly
(C)C-arm motion must be smooth
(D)Collimator light must be working”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

“21.What tool is used to evaluate for artifacts?
(A)Phantom
(B)SMPTE test pattern
(C)TG18 test pattern
(D)Monitor calibration tool”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

“A single binary digit of data, such as “0” or “1,” is called a
(A)pixel
(B)matrix
(C)bit
(D)byte

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

“23.In general, higher magnification will require the use of a
(A)larger focal spot size
(B)smaller OID
(C)smaller focal spot size
(D)larger SID

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

“If the magnification mammography is performed without using a small focal spot size, the resulting image will be magnified
(A)and blurred
(B)and sharply outlined
(C)with increased subject contrast
(D)with increased detail”

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

24.A grid is not necessary during magnification because
(A)grid use decreases spatial resolution
(B)the small focal spot size used will compensate for the loss of image detail
(C)the large OID produces the same effect as a grid
(D)magnification will magnify the normally invisible grid line”

A
25
Q

“25.The air gap in magnification allows increase subject contrast by
(A)increasing scatter
(B)reducing scatter
(C)reducing motion
(D)increasing motion”

A
26
Q

“At higher magnification factors there is
1.higher skin dose
2.increased scatter
3.decreased source-to-object distance
(A)1 and 2 only
(B)2 and 3 only
(C)1 and 3 only
(D)1, 2, and 3”

A
27
Q

“Magnification is beneficial in all of the following situations except
(A)imaging the surgical site of a patient undergoing a lumpectomy
(B)imaging a specimen radiograph
(C)evaluating microcalcifications in a lesion
(D)imaging the entire breast in a single exposure

A
28
Q

“Using a small focal spot size is recommended for magnification
(A)to compensate for the loss of image detail
(B)because of increased patient dose
(C)to compensate for the small OID
(D)to compensate for motion unsharpness”

A
29
Q

“Which of the following is a big disadvantage of magnification?
(A)Increased OID
(B)Increased patient dose
(C)Decreased subject contrast
(D)Increased risk of motion unsharpness”

A
30
Q

“If the backup time stops a breast exposure, the technologist can repeat the radiograph by using a
(A)higher peak kilovoltage (kVp) setting
(B)higher brightness compensation
(C)higher milliampere second (mAs) setting
(D)different automatic exposure control (AEC) setting”

A
31
Q

“The type of x-rays created from displacement of K-shell–binding electrons in the molybdenum atom are called
(A)coherent scattering
(B)characteristic radiation
(C)Compton effect
(D)bremsstrahlung radiation”

A
32
Q

“33.The function of the filter in mammography is to remove
(A)low-energy x-rays that would result in increased skin dose
(B)all low-energy photons
(C)low-energy x-rays needed to produce the breast image
(D)photon energies higher and lower than the range needed for breast imaging”

A
33
Q

“34.A recommended labeling for the mammographic image that is not required by the Mammography Quality Standards Act (MQSA) is
(A)technologist identification
(B)date of service
(C)technical factors
(D)name of patient”

A
34
Q

“Lack of breast compression is most likely to cause
(A)geometric unsharpness
(B)detector artifact
(C)motion unsharpness
(D)artifact”

A
35
Q

“How long should a facility maintain records of serious complaints?
(A)1 year
(B)2 years
(C)3 years
(D)4 years”

A
36
Q

“Which of the following focal spot sizes is used in routine imaging?
(A)0.3
(B)0.15
(C)0.5
(D)0.1”

A
37
Q

“Which of the following systems does not use a grid?
(A)All flat-panel detector tomosynthesis imaging systems
(B)2D mammography system
(C)Any magnification mammography system
(D)Indirect technology imaging systems”

A
38
Q

“Aluminum (Al) can be used as the filtration in
(A)digital breast tomosynthesis tubes with tungsten targets
(B)tungsten target digital mammography tubes
(C)rhodium target digital mammography tubes
(D)all mammography tubes”

A
39
Q

“The detector element (DEL) is located in the
(A)computer
(B)flat-panel detector
(C)thin-film transistor (TFT)
(D)AEC”

A
40
Q

“Which of the following statements about detector elements is incorrect?
(A)The detector element size helps to control the spatial resolution of the system
(B)The size of the detector element can be controlled by the technologist
(C)After reading, the detector element is automatically erased
(D)The larger the detector element in a flat-panel detector, the more the image blur”

A
41
Q

“To allow full coverage of the detector, a tube with a 16-degree anode angle requires a tube tilt of approximately _____ degrees to achieve an effective anode angle of 22 degrees.
(A)2
(B)4
(C)6
(D)8”

A
42
Q

“43.Which of the following monitors is used to interpret the mammograms?
(A)Acquisition workstation monitor
(B)Liquid-crystal display (LCD) monitor
(C)AWS monitor
(D)Review monitor”

A
43
Q

“Which of the following is a system that can be used to store 12 months or more of image data and can retrieve the data within a few minutes?
(A)Jukebox
(B)Redundant array of independent (inexpensive) discs (RAID)
(C)Digital linear tape (DLT)
(D)Compact disc (CD)”

A
44
Q

“A computer network system that allows images to be viewed at various monitors or transmitted or stored is termed
(A)picture archiving and communication system (PACS)
(B)local area network (LAN)
(C)wide area network (WAN)
(D)Digital Imaging and Communications in Medicine (DICOM)

A
45
Q

“The remote transmission of images is
(A)RAID
(B)DICOM
(C)PACS
(D)teleradiology”

A
46
Q

“What is the purpose of CAD?
(A)Tomographic image reconstruction
(B)3D image recording
(C)Use to provide final mammography interpretation
(D)A method of prereading mammograms

A
47
Q

“Which of the following is included in a workflow plan for a technologist?
(A)Determining whether a lesion requires biopsy
(B)Completed the medical audit
(C)Retrieving the results of prior examinations for the patient
(D)Monitoring the radiation output of the unit”

A
48
Q

“What system is identified as a storage for radiology imaging information?
(A)HIS
(B)DICOM
(C)RIS
(D)EMR”

A
49
Q

“When transmitting mammography images, the compression required is termed
(A)lossy compression
(B)PACS compression
(C)DICOM compression
(D)lossless compressing

A
50
Q

“The device used to convert electronic signals to computer language is called
(A)ADC
(B)DAC
(C)DBT
(D)PAC”

A
51
Q

“52.Changing the image contrast would mean adjusting the
(A)window width
(B)window level
(C)window length
(D)window height”

A
52
Q

“53.In direct digital technology, the x-ray strike _____ in the detector.
(A)amorphous selenium
(B)amorphous silicon
(C)cesium iodide
(D)scintillator

A
53
Q

54.What test tool is used to calibrate the brightness of the monitor?
(A)TG15
(B)SMPTE
(C)SEMPT
(D)AAPM”

A
54
Q

“55.Which of the following is a true statement according to the EQUIP mandated by FDA?
(A)Each breast center must have a written policy explaining the documentation of any corrective actions when images are of poor quality
(B)Documentation is not required for corrective action when images are of poor quality
(C)The policy for the lead interpreting physician (LIP) oversight of QA/QC records must be a written policy
(D)Written documentation is required for the clinical image review of each radiologic technologist (RT) and interpreting physician (IP), and it must be done at least annually”

A
55
Q

“What is the maximum possible score for each test object in the ACR phantom?
(A)4
(B)6
(C)2
(D)8

A
56
Q

“57.Synthesized 2D imaging is a
(A)two-dimensional (2D) reconstruction of the three-dimensional (3D) image
(B)3D reconstruction of the 2D image
(C)2D imaging with tomosynthesis
(D)3D tomosynthesis imaging”

A
57
Q

“Computer-aided detection (CAD) is used to
(A)visually enhance an image for viewing
(B)modify the contrast and brightness of an image
(C)analyze or preread the image
(D)provide accurate diagnostic information about the image”

A
58
Q

“What BI-RAD category recommends short-interval (6-month) follow-up?
(A)Category 2
(B)Category 3
(C)Category 4
(D)Category 5”

A
59
Q

“The patient has a BI-RADS category 5 report. This finding indicates
(A)a high possibility of malignancy
(B)a known malignancy
(C)a benign finding that needs a biopsy
(D)a suspicious abnormality”

A