Chapter 13 Flashcards
exposure factor
the factors that influence and determine the quantity and quality of x-radiation to which the patient is exposed
four prime exposure factors
kVP, mA, exposure time, SID
kVP and mAs
the most important factors principally responsible for x-ray quality and quantity. focal spot size, distance, and filtration are secondary factors that may require manipulation for particular examinations
kVP
affects quality and therefore, beam penetrability,with increasing kVP, more x-rays are emitted, and they have higher energy and greater penetrability. but because they have higher energy, they also interact more by Compton effect and produce more scatter radiation, which results in reduced image contrast.
the kVP selected helps to determines the number of x-rays in the image forming beam, and hence the resulting average optical density (OD).
finally, and most importantly, the kVP controls the scale of contrast on the finished radiograph because as kVp increases, less differential absorption occurs, therefore, high kVP results in reduced image contrast.
Milliamperes
the mA selected determines the number of x-rays produced and therefore the radiation quantity. the unit of electric current is the ampere (A) one ampere is equal to 1 coulomb (c) of electrostatic charge flowing each second in a conducter.
fallng load generator
on an x-ray imaging system in which only mAs can be selected, exposure factors are adjusted automatically to the highest mA at the shortest exposure time allowed by the high voltage generator
Distance
affects exposure of the image receptor according to the inverse square law, the SID largely determines the intensity of the x-ray beam at the image receptor.
direct square law
is derived from the inverse square law, it allows a radiologic technologists to calculate the required change in mAs after a change in SID to maintain constant OD.
focal spot size
for general imaging, the large focal spot is used. this ensures that sufficient mAs can be used to image thick or dense body parts. the large focal spot also provides for a shorter exposure time, which minimizes motion blur.
one difference between large and small focal spots is the capacity to produce x-rays. many more x-rays can be produced with the large focal spot because anode heat capacity is higher. with small focal spot, electron interaction occurs over a much smaller area of the anode, and the resulting heat limits the capacity of x-ray production.
a small focal spot is reserved for fine-detail radiography, in which the quantity of x-rays is relatively low. small focal spots are always used for magnification radiography. these are normally used during extremity radiography and in examination of other thin body parts in which higher x-ray quantity is not necessary.
filtration
three types of x-ray filtration are used: inherent, added, and compensating.
inherent filtration
all x-ray beams are affected by the inherent filtration properties of the glass or metal envelope of the x-ray tube. the value of the inherent filtration is approximately 0.5 mm the required total filtration of 2.5 mm is needed.
compensating filters
are shapes of aluminum mounted onto a transparent panel that slides in grooves beneath the collimator. these filters balance the intensity of the x-ray beam so as to deliver a more uniform exposure to the image receptor. they may be shaped like a wedge for examination of the spine or like a trough for chest examination.
as added filtration is increased, the result is increased x-ray beam quality and penetrability. the result on the image is the same as that for increased kVP, that is, more scatter radiation and reduced image contrast.
three basic types of high voltage generators are available
single phase, three phase, and high frequency
high wave rectified generator
has 100% voltage ripple. during exposure with a half wave rectified generator, x-rays are produced and emitted only half the time. during each negative half cycle, no x-rays are emitted.
full wave rectification
identical to half wave rectification except there is no dead time. during exposure, x-rays are emitted continually as pulses. consequently, the required exposure time for full wave rectification is only half that for half wave rectification.
three phase power
comes in two principal forms: 6 pulse or 12 pulse. the difference is determined by the manner in which the high voltage step up transformer is engineered.
the difference between the two forms is minor but does cause a detectable change in x-ray quantity and quality. three phase power is more efficient than single phase power. more x-rays are produced for a given mAs setting, and the average energy of those x-rays is higher. the x-radiation emitted is nearly constant rather than pulsed.
patient factors
such as anatomical thickness and body compositon
image quality factors
such as OD, contrast, detail, and distortion
exposure technique factors
such as kVP, milliamperage, exposure time, and SID, as well as grids, screens, focal spot size, and filtration.
body habitus
sthenic – meaning “strong, active” patients are average.
hyposthenic–are thin but healthy appearing; these patients require less radiographic technique.
hypersthenic—are big in frame, and usually overweight.
asthenic are small, frail, sometimes emaciated, and often elderly.
calipers
are used to measure the thickness of the anatomy that is being irradiated
composition
when only soft tissue is being imaged, low kVp and high mAs are used. with an extremity, which consists of soft tissue and bone, low kVp is used because the body part is thin.
when imaging the chest, the radiologic technologist takes advantage of the high subject contrast. lung tissue has very low mass density, bony structures have high mass density, andd the mediastinal structures have intermediate mass density. consequently, high kVp and low mAs can be used to good advantage. this results in an image with satisfactory contrast and low patient radiation dose.
radiolucent
attenuates few x-rays and appears black on the radiograph
radiopaque
tissue absorbs x-rays and appears white on the radiograph
pathology (destructive)
causing the tissue to be more radiolucent
pathology (constructively)
increase mass density or composition, causing the tissue to be more radiopaaque
image quality factor
refers to characteristics of the radiographic image; these include OD, contrast, image detail, and distortion. image quality factor are considered the “language” of radiography.
Optical density
is the degree of blackening of the finished radiography. OD has a numeric value and can be present in varying degrees, from completely black, in which no light is transmitted to almost clear. whereas black is numerically equivalent to an OD of 3 or greater, clear is less than 2 a radiograph that is too dark has a high OD caused by overexposure. this situation results when too much x-radiation reaches the image receptor. a radiiograph that is too light has been exposed to too little x-radiation, resulting in underexposure and low OD.
Optical density controller
two major factors mAs and SID
it can be affected by other factors, but the mAs becomes the factor of choice for its control. a change in mAs of approximately 30% is required to produce a visible change in OD. as a general rule, when only the mAs setting is changed, it should be halved or doubled. the simplest method used to increase or decrease OD is to increase or decrease the mAs.
contrast
the function of contrast in the image is to make anatomy more visible. contrast is the difference in OD between adjacent anatomical structures, or the variation in OD on a radiograph. Contrast is perhaps the most important factor in radiographic quality.
penetrability of the x-ray beam is controlled by
kVP
gray scale of contrast
refers to the range of ODs from the whitest to the blackest part of the radiograph
high contrast radiographs produce short gray scale.
they exhibit black to white in just a few apparent steps
low contrast radiographs produce long gray scale
appearance of many shades of gray.
high contrast
” a lot of contrast” or a “short scale of contrast” is obtained by using low kVp exposure techniques
low contrast
is the same as “long scale of contrast” and results from high kVp
detail
describes the sharpness of appearance of small structures on the radiograph. with adequate detail, even the smallest parts of the anatomy are visible.
image detail
evaluated by two means–recorded detail and visibility of image detail
sharpness of image detail
refers to the structural lines or borders of tissues in the image and the amount of blur of the image. factors that usually control the sharpness of image detail are the geometric factors —focal spot size, SID, and OID. To produce the sharpest image detail, one should use the smallest appropriate focal spot and the longest SID and place the anatomical part as close to the image receptor as possible.
visibility of image detail
describes the ability to see the detail on the radiograph and is best measured by contrast resolution.
distortion
the misrepresentation of object size and shape on the radiograph.
elongation
means that the anatomical part of interest appears bigger than normal
foreshortening
means that the anatomical part appears smaller than normal
4 image quality factors
Optical density
contrast
detail
distortion
Optical density
controlled by
mAs
influenced by
kVp distance thickness of part mass density development time or temperature image receptor speed collimation grid ratio
Contrast
controlled by
kVp
influenced by
mAs (toe, shoulder) development time or temperature image receptor used collimation grid ratio
Detail
controlled by
focal spot size
influenced by
SID
OID
Motion
All factors related to density and contrast
Distortion
controlled by
patient positioning
influenced by
Alignment of tube
anatomical part
image receptor
exposure technique charts
kVP
mA
exposure time
SID
types of grids
stationary
moving grid
parallel or non focused grid
lead lines run parallel to one another
used primarily in fluoroscopy and mobile imagine
focused grid
has lead lines that are angled to approximately match the angle of divergence of the primary beam
Advantage
allow more transmitted photon to reach the film than parallel grid
convergent point
if imaginary lines were drawn from each of the lead lines in a linear focused grid, these lines would meet to form an imaginary point.
convergent line
if convergent points were connected along the length of the grid they would form an imaginary li
focal distance
distance between the grid and the convergent line or point
focal range
focal range is the recommended range of SIDs that can be used with focused grid.
the convergent line or point always falls within the focal range.
grid cassette
an image receptor that has a grid permanently mounted to its front surface.
grid cap
a grid cap contains a permanently mounted grid and allows the image receptor to slide in behind it.
disadvantage of stationary grid
causes shadows of the grid lines on the image
moving or reciprocating grid
they are part of the bucky (potter bucky diaphragem)
located directly below the radiograhic table top just above the tray that hold the film
grid motion is controlled electrically by the exposure switch
grid moves back and forth in a lateral direction over the image receptor during the entire exposure.
contrast improvement factor
principal function of a grid is to improve contrast
grid conversion factor (bucky factor)
grid also absorb some of the primary radiation
to compensate for this you need to increase the mAs
GCF = mAs with grid/mAs without grid
bucky factor/grid conversion factor
grid ratio bucky factor/GCF
non grid 1
5: 1 2
6: 1 3
8: 1 4
12: 1 5
16: 1 6
grid cutoff
grid cutoff is defined as a decrease in the number of transmitted photons that reach the image receptor because of some misalignment of the grid.
higher grid ratio results in more grid cutoff.
high ratio grid has less positioning latitude
upside down focused grid
appears radiographically as significant loss of density along the edges of the image.
off level grid cutoff
caused from angling the x-ray tube across the grid lines or angling the grid itself during exposure
appears as an overall decrease in density
off center grid cutoff
also called lateral decentering.
occurs when the central ray of the x-ray beam is not aligned with the center of a focused grid (side to side misalignment)
appears as an overall loss of density
off focused grid cutoff
occurs when using an SID outside of the recommended focal range.
occurs if the SID is less than or greater than the focal range.
appears as a loss of density at the periphery of the film.
grid selection
below 90 kVp 8:1 grid is used
above 90 kVp grid ratio above 8:1 is used
grid selection factors
patient dose increase with increasing grid ratio
high ratio grids are used for high kVp examinations
patient dose at high kVp is less than that of low kVp
air gap technique
image receptor is 10 to 15 cm from the patient
alternate to grid
improve image contrast
10% increase in maS for every cm of gap.
image magnification with associated focal spot blur
film construction
radiographic film has many layers
- supercoat
- emulsion
- adhesive layer
- film base
supercoat (overcoat)
supercoat is a durable protective layer that is intended to prevent damage to the sensitive emulsion layer underneath it.
emulsion layer
emulsion layer is the radiation and light sensitive layer of the film.
the emulsion of the film consists of silver halide crystals suspended in gelatin
silver halide contains 90 to 99% of silver bromide acid 1% to 10% of silver iodide.
adhesive layer
adhesive layer keep the emulsion sticking to the film base.
film base
the final layer of the film is base
base is made of polyester (plastic) which gives the film physical stability
most film base has a blue tint.
historic development of film base
glass: break easily, difficult to store
cellulose nitrate: highly flammable
cellulose acetate: could damage when it is wet
polyester base: most modern. dimensional stability
characteristics to be considered when selecting radiographic film
contrast speed spectral matching anti-crossover layer requirement for safelight
contrast
contrast of an IR is inversely propertional to its exposure latitude
exposure latitude = range of exposure techniques that will produce an acceptable radiograph.
depends on size and distribution of the silver halide crystals.
low, medium, or high
screen film
screen films are available with many contrast levels.
high contrast emulsion: smaller silver halide grains with uniform grain size.
low contrast emulsion: larger grains having wide range of sizes.
speed
sensitivity of the screen film combination to x-rays and light
different film emulsion and different intensifying screen phosphors contribute to different speed.
for direct exposure film, speed is affected by the concentration and total number of silver halide crystals.
factors affecting screen film speed
silver halide grain size
shape of the grains
concentration of the grains
double emulsion films are twice as fast as single emulsion films
covering power of the emulsion
current emulsions contain less silver, yet produce the same density per unit exposure because of its increased covering power.
cross over effect
crossover refers to a phenomena in which light from one intensifying screen cross over the film base and expose the emulsion on the opposite side of the base.
it causes increased blur on the image
crossover is a problem that is unique to double emulsion film used with intensifying screen.
crossover can be reduced by
tabular grain emulsion
anti crossover layer