chapter 11 Flashcards
radiation generated in the x-ray tube that does not exit form the collimator opening but rather penetrates the protective tube house and to some degree the sides of the collimator
leakage radiation
radiation that emerges directly from the x-ray tube collimator and moves without deflection toward a wall, door, viewing window and so on. also called direct radiation or useful beam
primary beam
Tube housing is constructed so that leakage radiation measured at a distance of 1m from the x-ray source does no exceed _____ when the tube is operated at its highest voltage at the highest current that allows continuous
1mGya/hr (100mR/hr)
2 other functions of the tube housing
- confines the high voltage entering the tube to prevent electric shock
- makes cooling of the x-ray tube possible
what is the radiographic table made of
carbon fiber material
For the SID the distance and centering indicators must be accurate within what percent of the SID respectively
2% and 1%
collimation is accomplishes primarily through the use of what?
light localized, variable aperture collimator
what are other types of collimators
aperature diaphragms and cones and cylinders
what does the 2 sets of collimators do?
1st set- reduces the amount of off focus radiation
2nd set- consists of 2 pairs that can be adjusted independently
to minimize skin exposure to electrons produced by photon interaction with the collimator, the patients skin surface should be how far below the collimator
15cm
what do some collimators have to prevent collimators from being closer than 15 cm to the patient
spacer bars
what is luminance
refers to the brightness and quantifies the intensity of the light source
what is luminance measured in
nit or candela per square meter
the x-ray beam and the light beam must coincide within ___% of the SID
2%
consists of electronic sensors that sends signals to collimator housing. when activated collimators are automatically adjusted to radiation field matches the size of the IR
Positive beam limitation (PBL)
State regulatory standards require accuracy of __% of the SID with PBL.
2%, some states may require an accuracy of only 35
simplest of all beam limits devices, consists of a flat piece of lead with a hole of designated size and shape that is attached to the x-ray tube head
aperture diaphragm
what is the purpose of filtration
reduces exposure to patients skin and superficial tissue by absorbing most of the lower energy photons (long wave) from the heterogeneous beam, it increases the mean energy, or quality of the x-ray beam, hardening of the beam
the absorbed dose of the patient ____ when the correct amount of types of filtration are placed in the path of the radiographic beam
decrease, it absorbs some of the photons and decreased the quantity of incident radiation. the remaining photons in the beam as a whole, are more likely to be absorbed in body tissue
Two types of filtration
inherent filtration and added filtration
what are the parts of the inherent filtration
glass envelope, insulating oil, and glass window
what glass window inherent material amounts to ___ of aluminum equivalent
0.5mm Al
What can they use to added filtration
sheets of aluminum (or equivalent) of appropriate thickness.
the regulatory standard for total filtration of ____ is required for fixed units operating above 70kVp
2.5mm Al Eq
what determines the amount of filtration required
kVp
stationary fixed radiographic equipment operating at 50-70kVp require a total filtration of?
1.5mm Al Eq
stationary fixed radiographic equipment operating below 50 kVp requires a total filtration of?
0.5mm Al Eq
mobile diagnostic units and fluoro equipment require a minimum of?
2.5mm Al Eq
what are the commonly used filters in mammo
molybdenum and rhodium
when the x-ray target is made of molybdenum each following filters may be used?
0.3 mm molybdenum filter and 0.025mm rhodium filter