ch 36 p3 Flashcards
Which of the following small volume jet nebulizer design features affect its performance? A. position B. Residual volume C. Baffles D. Reservoirs
All items
All of the following are categories of nebulizers accept: A. pneumatic jet nebulizers B. USNs C. VM nebulizers D. DMI
D.
Small volume jet nebulizers are used to commonly disperse what volume of medications? A. 20 to 50 mg B. 5 to 20 ml C. 20 to 50 ml D. 5 to 20 mg
B. SVN most commonly used for medical aerosol therapy hold 5 to 20 ml of medication
How much medication can a large volume jet nebulizer hold? A. 500 ml B. 100 Ml C. 200 Ml D. 150 Ml
C. 200 ml
Large volume jet nebulizers can hold up to 200 ml of medication
All of the following drug formulation characteristics affect the performance of small volume jet nebulizers accept: A. homogeneity B. Surface tension C. Viscosity D. Potency
D. Potency of drug does not affect performance of SVN
What is the average amount of dead volume in a small volume jet nebulizer after the device runs dry? A. 0.1 ml B. 0.5 to 2.2 ml C. 2.0 to 4.0 ml D. less than 0.1 ml
B. the residual volume of a 3 ml dose varies from as little as 0.5 ml to more than 2.2 ml, which can be more than 2/3 of the total dose
During aerosol drug delivery using a SVN set at 8 L/min input flow, a patient asks that the head of the bed be lowered to semi fowler position, immediately after doing so, you observe a significant drop in SVN output despite the fact that there is at least 3 ml of solution left in reservoir, what would correct the problem?
A. add 1 to 2 ml more diluent to the reservoir
B. increase the nebulizer flow to 10 to 12 L/min
C. reposition the patient so that the SVN is more upright
D. Decrease the nebulizer input flow to 3 to 4 L/min
C. reposition the patient so that the SVN is more upright
The higher the flow of gas generated in a nebulizer, the ______ size of particle size.
A. Higher
B. Smaller
B. Smaller
Droplet size and nebulization time are inversely proportional to gas flow through the jet
What is the name for the dead volume of medication remaining after the device stops generating aerosol? A. Residual drug volume B. Viscosity C. Liquid in reservoir D. Drug loss
A. Residual drug volume
You increase the fill volume from 2 to 4 ml in a SVN being used to administer a bronchodilator agent with an aerosol. What effect will this have on the amount of drug delivered? A. No effect B. Increase C. Decrease D. More waste
B. Increase
Increasing the fill volume allows a greater proportion of active medication to be nebulized
What happens as the pressure or flow delivered through a small volume jet nebulizer gets higher? A. Treatment time becomes shorter B. Particle size becomes smaller C. Aerosol output becomes greater D. All of the above
D. aLL
The higher the pressure or flow, the smaller is the particle size, the greater is the output, and the shorter is the treatment time
Which of the following is the effect of aerosol particles entrained into a warm and fully saturated gas stream? A. No effect B. Increase in size C. Decrease in size D. Increase in number
B. Increase in size
Aerosol particles entrained into a warm and full saturated gas stream increase in size
What amount of medicine is wasted when using a continuous dose SVN? A. 50% B. 60% C. 40% D. 20%
B: 60%
What percentage of deposition is measured with the use of SVN? A. 50 B. 10% C. 15% D. 45%
B. 10%
What can conserve aerosolized medications with a SVN? A. Baffle B. Reservoir C. Nothing, you must move to LVN D. one way valve
B. Reservoir
Which of the following is false about the optimal technique for using a small volume jet nebulizer?
A. SVNS are less technique and device dependent
B. Slow inspiration improves SVN aerosol deposition
C. Deep breathing or breath holding improves SVN deposition
D. Use of a mouth piece or mask provides similar results
C. Use of an SVN is less technique and device dependent than use of a pressurized metered dose inhaler or dry powder inhaler delivery system. Slow inspiratory flow does optimize SVN aerosol deposition. Deep breathing & breath holding during SVN do little to enhance deposition. If patient is mouth breathing there is little difference in clinical response between therapy given by mouthpiece or mask
Normally, when using a 50 psi flowmeter to drive a small volume jet nebulizer, to what should you set the flow? A. 2 to 4 L/min B. 4 to 6 L/min C. 6 to 10 L/min D. 8 to 10 L/min
C. 6 to 10 L/min
To decrease the volume of dead space of a small volume jet nebulizer during drug administration what should you do?
A. Decrease nebulizer flow
B. Turn the nebulizer upside down
C. Continue treatment until nebulizer begins to sputter
D. Increase the nebulizer flow
C. Continue treatment until nebulizer begins to sputter
To minimize a patients infection risk between drug treatments with a small volume jet nebulizer what should you do?
A. rinse the svn with sterile water, air dry
B. Carefully repackage the SVN in its wet state
C. Rinse the SVN with tap water, run until dry
D. Throw out the SVN after every treatment
A. The CDC recommends they be cleaned and disinfected with sterile water and air dried between uses
A patient with an acute exacerbation of asthma is not responding to the standard dose and frequency of an aerosolized bronchodilator and is now receiving SVN therapy every 30 min. Which of the following would you recommend to the patients physician?
A. Discontinue the aerosolized bronchodilator
B. Increase the frequency of SVN to every 10 min
C. Consider continuous nebulization of the drug
D. Add more diluent to the SVN to extend time
C.
C. An alternative approach is to provide continuous nebulization with specialized large volume nebulizer
What is the major problem with using large volume nebulizers for continuous aerosol drug therapy? A. decreased pulmonary deposition B. drug reconcentration and toxicity C. Frequent interruption of therapy D. Greater waste of drug
B. A potential problem with continuous bronchodilator therapy is drug concentration increase
A physician ordered the antiviral agent ribavirin (Virazole) to be administered by aerosol to an infant with bronchiolitis. Which of the following devices would you recommend in this situation?
A. Hydrodynamic (babbington) nebulizer
B. Small particle aerosol generator (SPAG)
C. Ultrasonic (piezoelectric) nebulizer
D. Large volume heated nebulizer
B. SPAG small particle aersol generator
Was developed by ICN pharmaceuticals specifically for the administration of ribavirin to infants with respiratory syncytial virus infection
The small particle aerosol generator (SPAG) produces a small monodispersed aerosol through wich of the following?
A. Aerosol impaction by sequential baffling
B. Particle evaporation in a glass drying chamber
C. Use of an inert liquefied gas propellant
D. Aerosol generation using vibrational energy
B. Particle evaporation in a glass drying chamber
WHen using a small particle aerosol generator to administer ribavirin to an infant which pair of flow settings would be correct? Nebulizer Drying Chamber A. 8 L/min 8 L/min B. 5 L/min 5 L/min C. 8 L/min 4 L/min D. 4 L/min 8 L/min
A. Nebulizer flow should be maintained at approximately 7 L/min with total flow from both flowmeters no lower than 15 L/min
What serious problems are associated with the delivery of ribavirin using small particle aerosol generators?
A. Caregiver exposure to the drug aerosol
B. Drug reconcentration in the drying chamber
C. Drug precipitation in ventilator circuits?
D. All of the above
A & C
Drug precipitation can jam breathing valves or occlude ventilator circuits
The problem of ribavirin aerosol precipitation causing malfunction of ventilator circuits can be overcome by which of the following?
A. Placing a one way valve between the small particle aerosol generator (SPAG) and the circuit
B. Placing a HEPA filter proximal to exhalation valve
C. Decreasing the SPAGs total flow to below 10 L/min
A & B
The problem can be overcome by placing a one way valve between the SPAG and the circuit & filtering out the excess particles before they reach exhalation valve, changing filters frequently