Ch. 10 Test, Assessment of Respiratory Function Flashcards
- A pulse oximeter differentiates oxyhemoglobin from deoxygenated hemoglobin by which of the following methods?
a. Relating cyclical changes in light transmission through the sampling site.
b. Shining and comparing two wavelengths of light through the sampling site.
c. Direct measurement by a heated polarographic electrode applied to the skin.
d. Using a color sensing device that absorbs one wavelength of light through the skin.
ANS: B
A pulse oximeter uses spectrophotometry to differentiate between oxyhemoglobin and deoxygenated hemoglobin. Two wavelengths of light (660 and 940 nm) are shined through a sample site. Oxyhemoglobin absorbs more light at 940 nm (infrared [IR] light) than does deoxygenated hemoglobin. Deoxyhemoglobin absorbs more light at 660 nm. The use of cyclical changes in light transmission measured at the sampling site is the method to determine the pulse rate with a pulse oximeter. A heated polarographic electrode is used for transcutaneous partial pressure of oxygen (PtcO2) measurements. A color sensing device is used to detect the amount of carbon dioxide in exhaled gas.
An arterial blood gas should be done to confirm pulse oximetry findings less than a minimum of _____________.
a. 60%
b. 70%
c. 80%
d. 90%
ANS: C
When a patient’s oxygen saturation measured by pulse oximeter (SpO2) is less than 80% an arterial blood gas should be drawn to confirm the patient’s oxygenation status because a pulse oximeter is generally accurate for oxygen saturations greater than 80%.
A pulse oximeter reading will be most accurate when used with a patient in which of the following situations?
a. An intensive care unit patient with hyperbilirubinemia
b. A hypotensive patient receiving peripheral vasoconstrictors
c. An emergency department patient with evidence of smoke inhalation
d. An open heart patient receiving extracorporeal membrane oxygenation
ANS: A
Hyperbilirubinemia does not appear to affect pulse oximetry measurements as do low perfusion states, which are caused by the use of peripheral vasoconstrictors or extracorporeal membrane oxygenation (ECMO). Carbon monoxide poisoning will lead to an overestimation of oxygen saturation measured by pulse oximeter (SpO2).
A patient arrives in the emergency department via ambulance following rescue from a house fire. The instrument that would be most appropriate to assist the respiratory therapist in assessing this patient’s oxygenation status is which of the following?
a. Capnograph
b. Pulse oximeter
c. Calorimeter
d. CO-oximeter
ANS: D
Laboratory CO-oximeters measure four types of hemoglobin: oxyhemoglobin (O2Hb), deoxygenated hemoglobin (HHb), carboxyhemoglobin (COHb), and methemoglobin (MetHb). This is beneficial for patients who are suffering from smoke inhalation. The CO-oximeter provides the actual O2Hb and the COHb. Carbon monoxide produces an erroneously high oxygen saturation measured by pulse oximeter (SpO2). Therefore, if smoke inhalation is suspected, a CO-oximeter should be used to evaluate the oxygen saturation. Capnography is the measurement of carbon dioxide concentrations in exhaled gases and is used to assess proper airway placement. Calorimetry allows the clinician to estimate energy expenditure from measurements of oxygen consumption (O2) and carbon dioxide production (CO2). This measurement may be useful when weaning a patient from mechanical ventilation.
While trying to use a finger probe to assess a patient’s oxygenation status, the respiratory therapist finds that the pulse rate and the ECG monitor heart rate are not consistent and the oxygen saturation measured by pulse oximeter (SpO2) reading is blank. The patient is awake, alert, and in no obvious respiratory distress. The respiratory therapist should first take which of the following actions?
- Change the probe site.
- Draw an arterial blood gas.
- Adjust the probe position on the finger.
- Remove the probe, and perform a capillary refill test.
a. 1 and 2 only
b. 2 and 3 only
c. 3 and 4 only
d. 1 and 4 only
ANS: C
The fact that the patient is awake, alert, and in no respiratory distress decreases the likelihood that the problem is with the patient. Therefore, the first action in this case should not be to draw an arterial blood gas (ABG). In cases where the pulse oximeter cannot identify a pulsatile signal, the oxygen saturation measured by pulse oximeter (SpO2) reading may not be present. This could be alleviated by adjusting the probe position on the finger. Absent SpO2 readings could also be due to low perfusion states. Performing a capillary refill test on the finger being used for the probe would show whether or not the finger has adequate blood flow. If this is true, the next step would be to change the site.
Pulse oximetry is most useful in which of the following situations?
a. Determining when to extubate an adult
b. Prescribing oxygen therapy for neonates
c. Monitoring patients undergoing chest physical therapy
d. Establishing initial oxygen necessity for home care patients
ANS: C
The oxygen status of a patient being considered for extubation needs to be assessed by an arterial blood gas, not by pulse oximetry, because not only does the patient’s oxygen status need assessment, but the acid-base balance does as well. Pulse oximetry is not used as a basis for prescribing oxygen therapy in neonates. Neonatologists prefer to base oxygen therapy decisions on arterial partial pressure of oxygen (PaO2) rather than oxygen saturation. Pulse oximetry is useful for monitoring the oxygen status of patients undergoing chest physical therapy because it gives immediate results and is used for continuous monitoring. Pulse oximetry may not be as useful in prescribing oxygen therapy for home care patients.
The respiratory therapist has just stopped postural drainage for a 24-year-old patient with cystic fibrosis because of shortness of breath and slight cyanosis in the “head-down” position. The respiratory therapist should recommend which of the following adjustments to therapy?
a. Continue postural drainage and monitor patient with capnography.
b. Use only upright or flat postural drainage positions and draw an arterial blood gas (ABG).
c. Administer oxygen via nasal cannula and monitor with pulse oximetry.
d. Use a transcutaneous partial pressure of oxygen (PtcO2) monitor to assess the extent of hypoxemia.
ANS: C
The presence of slight cyanosis and shortness of breath in the “head-down” position is indicative of hypoxemia. The respiratory therapist should administer supplemental oxygen via nasal cannula (possibly 1-2 L/min) and monitor the patient with a continuous pulse oximetry. Capnography is not useful in detecting hypoxemia. Using flat postural drainage positions where the head is not lower than the shoulders has not been proven to be effective.
A patient receiving mechanical ventilation is being continuously monitored for oxygen saturation measured by pulse oximeter (SpO2) for the past 48 hours. When initially applied, the SpO2 and the arterial oxygen saturation (SaO2), as well as the pulse on the pulse oximeter, ECG, and manual pulse, were consistent. During clinical rounds, the respiratory therapist notices that although the probe is appropriately placed and capillary refill is normal, the SpO2 reading is down to 90% from 95%. The most appropriate immediate action is to do which of the following?
a. Replace the probe.
b. Reposition the patient.
c. Draw an arterial blood gas.
d. Move the probe to a different site.
ANS: C
The oxygen saturation measured by pulse oximeter (SpO2) has dropped from 95% to 90%. Since the SpO2 and arterial oxygen saturation (SaO2) previously correlated, this situation could mean that the patient is becoming hypoxemic. The probe is appropriately placed, so changing sites is not appropriate. The patient has already been checked for and has adequate circulation to the site of the probe, so moving the probe to a site with more perfusion is not appropriate. Therefore, the patient needs to have an arterial blood gas to ascertain the SaO2 and partial pressure of oxygen (PaO2).
A patient in the intensive care unit is receiving mechanical ventilation, has a pulmonary artery catheter in place, and is being monitored continuously with a capnometer. The patient’s arterial partial pressure of carbon dioxide (PaCO2) is 41 mm Hg and the partial pressure of end-tidal carbon dioxide (PetCO2) is 36 mm Hg. There is a sudden decrease in the PetCO2 to 18 mm Hg causing an alarm to sound. The most likely cause of this development is which of the following?
a. Hypovolemia
b. Apneic episode
c. Pulmonary embolism
d. Increased cardiac output
ANS: C
Pulmonary embolism will cause a decrease in blood flow to the lungs. This increases alveolar dead space and leads to a decrease in the partial pressure of end-tidal carbon dioxide (PetCO2). Hypovolemia would also cause a decrease in the PetCO2, but it would not occur as suddenly as it did in this situation. The fact that the patient has an indwelling pulmonary artery catheter increases the risk of developing a pulmonary embolism, which often will have a quick onset. An apneic episode would have increased the PetCO2. An increased cardiac output would increase the PetCO2 because increases in cardiac output result in better perfusion of the alveoli and a rise in PetCO2.
The capnogram in the figure is indicative of which of the following conditions?
a. Chronic obstructive pulmonary disease (COPD)
b. Cardiac arrest
c. Hyperventilation
d. Pulmonary embolism
ANS: A
Phase 3 becomes indistinguishable when physiological dead space increases, as in chronic obstructive pulmonary disease (COPD), and causes the capnogram to appear as it does in the figure. A cardiac arrest would lower the graph line to zero. Hyperventilation will decrease the alveolar plateau, but its shape would remain the same as a normal capnogram. A pulmonary embolism would also cause a drop in the alveolar plateau.
For a given minute ventilation, partial pressure of end-tidal carbon dioxide (PetCO2) is a function of which of the following?
- Metabolic rate
- Cardiac output
- Alveolar dead space
- Physiologic shunt
a. 1 and 3 only
b. 1, 2, and 3 only
c. 2, 3, and 4 only
d. 1, 2, 3, and 4
ANS: B
Changes in metabolic rate cause changes in partial pressure of end-tidal carbon dioxide (PetCO2). For instance, fever and shivering increase the metabolism and increase the PetCO2. A change in cardiac output will change PetCO2 because the heart transports the blood that carries the carbon dioxide (CO2) to the lungs for elimination. Increases in cardiac output will increase PetCO2. A change in dead space ventilation will also cause changes in the PetCO2. Increasing dead space will decrease the PetCO2.
The normal range for arterial-to-end-tidal partial pressure of carbon dioxide [P(a-et)CO2] is which of the following?
a. 2-4 mm Hg
b. 4-6 mm Hg
c. 6 -8 mm Hg
d. 8-10 mm Hg
ANS: B
The arterial-to-end-tidal partial pressure of carbon dioxide [P(a-et)CO2] for tidal breathing should be approximately 4-6 mm Hg.
During shift report, the day shift respiratory therapist informs the night shift respiratory therapist about a freshly postoperative patient who is receiving full support via mechanical ventilation. At the time of the last patient-ventilator system check the patient had not awaken from anesthesia. During first round on the day shift the respiratory therapist notes the capnography shown in the figure. The most appropriate action to take would be to do which of the following?
a. Administer a bronchodilator.
b. Begin the weaning process.
c. Fix the leak in the sampling line.
d. Reinflate the ET tube cuff.
ANS: B
The figure shows a patient whose capnography is demonstrating spontaneous respiratory efforts during mechanical ventilation. This corresponds to the patient’s waking up from anesthesia. With all else stable, the next step would be to begin the weaning process.
The area under the curve of a single-breath carbon dioxide (SBCO2) curve represents which of the following?
a. Tidal volume
b. Alveolar dead space
c. Physiologic dead space
d. Effective alveolar ventilation
ANS: D
The area under the single-breath carbon dioxide (SBCO2) curve represents alveolar volume. This is known as phase 3 on the SBCO2 curve. Phase 1 is the anatomical dead space volume. Phase 2 is a transitional phase between anatomical dead space and alveolar volume.
The change in the single-breath carbon dioxide (CO2) curve from “A” to “B” shown in the figure may be a result of which of the following?
a. Hypervolemia
b. Decreased positive end-expiratory pressure (PEEP)
c. Increased mean airway pressure
d. Excessive bronchodilator administration
ANS: C
The change in the figure shows an increase in phase 1 and a decrease in phase 2 of the single-breath carbon dioxide (SBCO2) curve. An increase in phase 1 may be caused by an increase in anatomical dead space. This is possible as a result of increased airway obstruction or excessive positive end-expiratory pressure (PEEP). Therefore, answers “B” and “D” cannot be correct. A decrease in phase 2 means there is a decrease in venous return or an increase in intrathoracic pressure. This could be caused by hypovolemia or an increased intrathoracic pressure as seen with increased mean airway pressures. Therefore, answer “A” cannot be correct. Answer “C” is a measure of increased intrathoracic pressure and is the correct answer.
A patient is receiving mechanical ventilation with a fractional inspired oxygen (FIO2) of 0.85 and a positive end-expiratory pressure (PEEP) of 5 cm H2O. His arterial partial pressure of oxygen (PaO2) is 68 mm Hg, arterial oxygen saturation (SaO2) is 88%, and partial pressure of end-tidal carbon dioxide (PetCO2) is 32 mm Hg. Over the next few minutes his PEEP is titrated resulting in the following data:
Time. 0600. 0630. 0650. 0720. 0740.
FIO2 0.85. 0.85. 0.85. 0.80. 0.80
PEEP (cm H2O). 5. 8. 10. 12. 15
SpO2 (%). 88. 88. 90. 93. 90
PetCO2 (mm Hg) 30 30 32. 34. 25
At 0740 the single-breath carbon dioxide (SBCO2) curve shifted to the right. What action should the respiratory therapist take at this time?
a. Increase the FIO2 to 0.90.
b. Reduce the set tidal volume.
c. Continue to increase the PEEP.
d. Reduce the PEEP to 12 cm H2O.
ANS: D
The increase in positive end-expiratory pressure (PEEP) to 15 cm H2O seems to have decreased pulmonary perfusion because of overinflation of the alveoli. This is evident by the decrease in the partial pressure of end-tidal carbon dioxide (PetCO2) to 25 mm Hg and the right shift in the single-breath carbon dioxide (SBCO2) curve. Increasing the fractional inspired oxygen (FIO2) will not address this problem. Reducing the set tidal volume will increase the PetCO2 but will not improve the pulmonary circulation. Continuing to increase the PEEP will further reduce pulmonary perfusion and cause more dead space. Reducing the PEEP back to 12 cm H2O will optimize the PEEP and reduce overinflation.