Capnography Flashcards
Capnography
Definition, what reflects
- Aka infrared spectography
- A rapid and accurate method to measure exhaled CO2 throughout the respiratory cycle.
- Typically measured at the junction between the patient and the breathing system
- Directly shows elimination of CO2 by the lungs
- Indirectly reflects elimination of CO2 by tissues and adequate perfusion of lungs
- etCO2 is the CO2 at the end of expiration
etCO2 compared to arterial CO2 in healthy adults with normal lungs
etCO2 is 0.3-0.6kPa less than the arterial CO2
Capnograph vs capnogram vs capnometer
Capnograph = records and shows the graphical display of waveform of CO2. Displays value of etCO2
Capnogram = graphical plot of CO2 partial pressure (or percentage) vs time (see picture)
Capnometer = instrument which only shows numerical concentration of CO2 (no waveform)
Capnography mechanism: absorption of radiation by CO2
- Capnograph works via infrared spectrometry
- Gases with molecules that contain at least 2 dissimilar atoms (eg. carbon and oxygen in CO2) absorb radiation in the **infrared region of the spectrum **
- CO2 absorbs radiation at wavelength of 4.3 micrometers
- Beer-Lambert law: the absorption of light depends on the properties of the material through which the light is travelling -> amount of radiation absorbed is proportional to the number of CO2 molecules (partial pressure of CO2)
Components of a capnograph (4)
Light source
* Infrared radiation emitted by a hot wire or light source
* Often microprocessor controlled to produce stable source with constant output
Interference filter
* Particular frequency of radiation required is obtained by passing radiation through an interference filter
* An optical filter in which the wavelengths not transmitted are removed by interference phenomena (i.e. rather than absorption or scattering)
Sample chamber and reference chamber
* Windows made from material transparent to infrared radiation e.g. sodium chloride, silver bromide or sapphire (because glass absorbs infrared radiation)
* Infrared passes simultaneously through sample chamber (containing sample gas) and reference chamber (containing room air). Absorption from each chamber is compared -> CO2 values calculated
Photodetector: thermopile detector or sensor
* After passing through the sample and reference chambers, radiation is focused on a photodetector
* Greater absorption of infrared radiation by gas in sample chamber -> less radiation falls onto detector
* Thermopile detector produces heat which is proportional to partial pressure of CO2 in the mixture in the sample chamber. Heat is measured by a temperature sensor
* -> changes in electrical output and voltage in a circuit
Note: due to large amount of infrared absorption by CO2 in the sample chamber, little infrared finally reaches the detector
Other methods to measure CO2 (4)
- Raman spectrometry
- Mass spectrometry
- Photo acoustic spectrometry
- Chemical colorimetric analysis
Raman spectrometry
Alternative method to measure CO2
* Gas exposed to monochromatic light from an argon laser
* Energy from the light is absorbed by the intermolecular bonds and partially re-emitted at new wavelengths by the molecules
* Wavelength shift and scattering can be used to measure hte concentration of gases in the system
Advantages:
* allows measurement of all the gases in breathing system (CO2, O2, N2O, volatile anaethetic agents)
* Small and portable
* Gases returned to breathing system unchanged
Disadvantages: very slow response rate
Mass spectrometry
Alternative method to measure CO2
* Gas drawn into the spectrometer where it is ionised, then exposed to magnetic field in a vacuum chamber
* Various gases are separated according to their mass to charge ratio
* Concentration of various gases at known mass to charge ratio then calculated
Advantages:
* Highly accurate, reproducible
* Measurement of many gases
Disadvantages:
* Equipment is very bulky and expensive
* Susceptible to damage from water and some drugs
* Ionized gases cannot be returned to the breathing system, must be scavenged
Photo acoustic spectrometry
Alternative method to measure CO2
* Sample gas is irradiated with a suitable wavelength pulsatile infrared radiation
* -> periodic expansion and contraction -> pressure fluctuation of aurible frequency that can be detected by a microphone in proportion to the amount of gas present
Advantages:
* Extremely stable, calibration remains constant over much longer periods of time compared to conventional infrared absorption spectrometry
* V fast rise and fall times -> more accurate representation of any change in CO2 concentration
Chemical colorimetric analysis
Alternative method to measure CO2
* CO2 hydrated -> carbonic acid -> can be measured by pH sensitive means
* pH sensitive dye undergoes a colour change in the presence of CO2. Usually metacresol purple, changes to yellow in presence of CO2
Advantages:
* Detectors are small and portable
* Sometimes used in the field/pre-hospital
Disadvantages:
* Only gives crude assessment of CO2 level (low, normal, high)
Capnograph response time
Transit time, rise time
To be clinically useful, capnograph needs rapid response time, which has two components: transit time and rise time.
Transit time: time for sample to be transferred to analyser.
* Depends on length (i.e. short as possible e.g. 0.2m) and diameter of the sampling tubing, and sampling rate
* Delay of <3.8s is acceptable
Rise time: time for analyzer to respond to the signal. Depends on size of sample chamber and the gas flow
Side stream capnograph: set-up
- Sampling chamber connected to distal end of breathing system via sampling tube
- Sampling tube is 1.2mm internal diameter tube that samples both inhaled and exhaled gases at constant rate of ~150-200ml/min. Made of teflon (so impermeable to carbon dioxide, does not react with anaesthetic agents)
- Light weight adaptor connects sampling tube to breathing system with small increase in the dead space.
- Moisture trap with an exhaust port allows gas to be vented to the atmosphere or returned to the breathing system
Key points for use
* Sampling tube should be placed close to patients trachea for accuracy
* Other gases and vapours CAN also be analyzed from the same sample
Main stream capnograph: set-up
- Sampling chamber positioned within patient’s gas stream -> increasing the dead space
- Heated to about 41 degrees C to prevent water vapour condensation on its windows
Key points
* No sampling tube: no transport time delay in gas delivery to sample chamber
* CANNOT measure other gases and vapours simultaneously
Side stream vs main stream capnography: advantages
Side stream capnograph:
* Easy to connect
* Can be used in awake or non-intubated patients
* Easy to use when patient is in unusual positions e.g prone positioning
* No problems with sterilisation
* Multiple gas analysis (but calibration gas needed)
Main stream capnograph:
* Suitable for neonates and children
* No sampling tube -> no transport time delay
* No effect due to pressure drop
* No effect due to changes in water vapour pressure
Note, both produce a capnogram
Label the phases of this normal waveform
- I, baseline: Inhalation ends, exhalation begins, dead space air is eliminated first and no CO2 is present
- II, ascending phase, early exhalation: alveolar gas begins to mix with dead space air, sharp upstroke is produced
- III, alveolar plateau: alveolar air predominates, CO2 level plateaus. etCO2 is noted at end of exhalation (D)
- IV, descending phase, inhalation: inhalation occurs, CO2 level returns to baseline