C. dose calculation Flashcards
The beam energy that is used for treatment is determined by the
depth of the target volume, minimum target dose, and acceptable dose to organs at risk in path of beam
For dosimetric calculations, all rectangular treatment fields must be converted to
squares
What is equivalent square equation?
EQSQ= (2ab) / (a+B) or ( 4 x area)/ perimeter
_____fields are fields that use MLCs or blocks to change the shape of the treatment field
Blocked
A simpler approach is to take the square root of the open area minus the
blocked areaExample: the open area is 20 x 10 and the blocked area is 4x 6 First do the equation (20 x 10) — (4 x 6) Next, take the square root of the above answerClarkson integration can be used to find the area of irregularly shaped fields
The main cause of collimator scatter (sc) is from photons scattering within the
head of the gantry
Collimator scatter (Sc) Mostly occurs with the
collimating jaws and the flattening filter
Bigger field sizes creates ___ scatter
more
Smaller field size creates ___ scatter
less
Scatter that happens within the patient or phantom is called
phantom scatter (Sp)
phantom scatter (Sp) amount depends on the
volume of tissue within the treatment field
As the field size increases, the amount of phantom scatter ___
increases
Backscatter factor (BSF) and peak scatter factor (PSF) both compare the dose rate in free space to the dose rate within
a phantom at the dmax
Backscatter factor (BSF) is used for ___ beams
low-energy beams
Peak scatter factor (PSF) is used for ____ beams
high-energy beams
____ is the depth where electronic equilibrium is reached and 100 percent of the dose is delivered
dmax
The dmax for higher-energy beams is
deeper
What is dmax for 4MV beam
1 cm
What is dmax for 6 MV beam
1.5cm
What is dmax for 10MV beam
2.5 cm
What is dmax for 15MV beam
3 cm
What is dmax for 18MV beam
3.2cm
What is dmax for 20MV beam
3.5 cm
What is dmax for 25MV beam
5 cm
What is dmax for 6MeV beam
1.5cm
What is dmax for 9MeV beam
2.2cm
What is dmax for 12MeV beam
2.8cm
What is dmax for co-60
0.5cm
_____ is A ratio that compares the absorbed dose at a depth to the absorbed dose at a specific reference depth (dmax)
PDD= percent depth dose
What is equation for PDD?
[(Absorbed dose at depth) ÷ (absorbed dose at dmax)] x 100
PDD i only used to calculate MU in
SSD treatments
PDD changes depending on four components
energy, depth, field size, and SSD
An increase of ____ causes an increase of PDD because there is more scatter, which contributes to the dose
field size
As depth within the patient goes past Dmax then PDD
decreases
PDD and depth have a
inverse relationship