Breast MQSA Flashcards
QA Frequency: Processor QC
Daily
QA Frequency: Darkroom cleanliness
Daily
QA Frequency: Screen-Film contrast
Semi-Annually
QA Frequency: Viewbox conditions
Weekly
QA Frequency: Darkroom Fog
Semi-Annually
QA Frequency: Repeat Analysis
Quarterly
QA Frequency: Phantom Evaluation
Weekly
QA Frequency: Compression Test
Semi-Annually
Recall %
5-7% (<10%)
Cancers / 1000 screen
3-8
Resident Required months
3 total
Resident’s required mammograms in the last two years, within 6 months
240
Documented education hours
60
How long to send Lay report
30 days
Who’s responsible for QC program?
Interpreting Physician
Required resolution A-C and L-R directions
AC: 13 lp/mm and L-R: 11lp/mm
Image QC phantoms
4 fibers
3 microcalcifications
3 masses
“acceptable artifacts”
Dose phantom
50 % glandularity
4.2 cm thick
<3mGy per image
Patient dose?
No patient dose. Phantom dose only.
Typical dose per view?
2 mGy view, 8mGy exam
Why does digital have lower dose?
Higher beam quality