Brachytherapy Flashcards
Give some examples of interlocks that need to be checked in an HDR suite
Guide tube length ALMO radiation monitor Interrupt buttons Emergency stop Door interlock Battery backup
What does RAKR stand for
Reference Air Kerma Rate
Define RAKR
Reference Air Kerma Rate
The kerma rate to air, in air, at a reference distance of 1m, corrected for air attenuation and scattering
Where is RAKR defined in the literature?
IPEM COP for HDR BT
Why is using a well chamber better than using a thimble chamber for calibration?
Less sensitive to source positioning errors due to 4pi geometry.
Much smaller uncertainty
Hard to get reproducible positions when using a thimble
Much quicker to take measurements with a well chamber
Transfer uncertainty between well chambers smaller than between Farmer chambers
What is the primary standard for RAKR at the NPL?
Graphite-walled cavity ionisation chamber
What are the two steps for the calibration of secondary standard chambers?
RAKR of Ir-192 source determined using NPL primary standard
Calibrated source used to calibrated SS instrument
What is the source calibration set-up used at NPL?
Source inside a lead collimator with aperture such that the direction from the source centre to the reference point is at right angles to the long axis of the source.
Collimated beam directed towards primary standard
Centre-to-centre source-to-chamber distance 1433mm
Why are the source-to-chamber distance and aperture sizes chosen as they are at the NPL calibration?
To give a uniform field over the whole of the ionisation chamber
What correction factors are applied to determine the RAKR
Temperature and pressure Normalised to 1m Air attenuation Scatter correction factor Primary standard sensitivity
Reference time and date
What is the secondary standard calibration set-up that should be used?
Well chamber 1m from any wall and 1m above floor level on a low scatter surface.
Time for chamber to reach thermal equilibrium.
Bias voltage s.t. central collecting electrode is positive wrt outer electrode.
What is the position of the source that is used in the chamber for calibration?
The “sweet spot”, or position of maximum response .
What is the equation for RAKR in terms of Gy/s at 1m?
Kdot_R = M x k_ion x k_sg x N_Kdot_R
What correction factors are applied to the instrument response?
Electrometer correction factor
Decay correction factor to correct the measured current to a reference time
Temperature and pressure
At what threshold should discrepancies between the manufacturer’s source certificate and the measurement be investigated?
3%
At what threshold discrepancy between calculation and certificate should a source not be used clinically?
5%