Device alerts - Shock/ATP

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Device alerts - HM disconnected

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Device alerts - Lead integrity

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

ICD

Define detection window

A

Uses a moving window of 10 intervals to make decisions

Detection is satisfied when 8/10 intervals are fast

Rate is the criterion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

ICD

Define duration

A

Durations ensures that the arrhythmia is sustained before initiating treatment

6/10 must remain fast

Rate and duration are criterion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What determines which therapy is delivered?

A

Last interval in the zone

*last interval must be fast and in the zone for therapies to occur

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Identify BS detection enhancements

A

Onset
Stability
V rate > A rate
A fib rate threshold
Sustained Rate duration
Morphology (Rhythm ID)

*only enhancements programmed ON at GOSH are Rhythm ID in VT and VT mom zone.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

How many stages does BS reconfirmation have?

A

2 stages:
- during charge: 6/10
- After charge: 2/3

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Programming for history of VF/SCD

A

VF - 180/230bpm - ATP, then shocks
VT - 160bpm - monitor only

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Programming for history of VT/VF

A

VF - 200bpm - ATP, then shocks
VT - 160bpm - ATP, the shocks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Programming prophylaxis for VT/VF

A

VF - 180bpm - ATP, the shocks
VT - 160bpm - Monitor only

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Programming for VF only (Long QT etc.)

A

VF - 220bpm - shocks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What Detection enhancements are only found in single chamber BS devices?

A

Onset
Stability
Shock if unstable
Sustained rate discrimination (SRD)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What is the magnet rate for Boston devices at each stage of life?

A
  1. BOL = 100
  2. ERI = 90
  3. EOL = 85
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

For Emblem SICD and Embelm SICD MRI, what models are MRI compatible?

A

A209
A219

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Which Boston SCID leads are considered MRI compatible?

A

Boston Scientific Emblem SICD electrode (3401 & 3501)

Cameron health Q-TRACK SICD electrode (3010)

17
Q

What is the recommended magnet strength for MRI scans?

A

1.5 T

18
Q

Identify 3 contraindications for MRI scan in cardiac devices

A
  1. Non MRI compatible system
  2. Abandoned devices, components or accessories
  3. <6 weeks post implant/intervention
19
Q

What features and functions are suspended during MRI mode?

A
  1. Tachy detection & therapy
  2. System diagnostics
  3. Magnet detection
20
Q

Purpose of ventricular tachy-arrhythmia therapy

A

Treat arrhythmias associated with sudden cardiac death - VT/VF

21
Q

Purpose of bradycardia pacing

A

To detect and treat bradyarrythmias and provide cardiac rate support after defibrillation therapy

22
Q

Identify the shock vectors in Boston devices

A
  1. Distal shock electrode to proximal shock electrode & can
  2. RV to RA coil
  3. RV coil to can
23
Q

Boston scientific pulse generators are contraindicated in which patients?

A
  1. Digitalis
  2. Electrolyte imbalance
  3. Hypoxia
  4. Sepsis

All reversible causes of VF