Block 1 Flashcards
The basic framework consists of a collimated detector system capable of acquiring data at varying degrees within a
360° Radius, ADC conversion
Choice of collimator changes the
Image resolution
Does spect improve sensitivity or specificity
Sensitivity and specificity
Which two areas are CT used
SPECT- Thallium etc. & PET - Rubidium etc.
One limitation of SPECT/scintillation cam
It depicts 3D images in 2D until after reconstruction
SPECT Removes
Out of plane information
Which collimator is used for SPECT
Parallel hole
What focal slices are done with SPECT
Transverse, coronal, sagital
SPECT stands for
Single photon emission computed tomography
Two limitations to SPECT
Limited number of projection angles and the in-focus plane image still is obscured by counts from the former
Does the parallel.hole collimator change image size?
No it stays the same, no mag or min
Transverse is also called
Short axis
What increments are the data stored in
64 x 64
128 x 128
256 x 256 matrix
Describe each image plane
Coronal- front to back
Saggital- left and right
Transverse- superior and inferior
Most popular way to do SPECT
With a rotating Anger/angle camera mounted on a gantry that allows 360° rotation
Most common detector system
Two or three heads which increase sensitivity
T or F: a detector acquires data from a thin axial section containing radioactivity by linear scanning from multiple directions around the cross section
True
What is a projection
The counts of the detector at different positions along a scan line at different angles
Multiplying for matrix size example
Example: 8x8 is 64
So 64x64 is 4096
128x128 is 16384
256x256 is 65536
Pixel width determines the
Resolution
Reducing pixel width improves
Resolution
If you reduce pixel width…
Resolution increases, sensitivity decreases
Cardiac SPECT looks at which area of the heart
Left ventricle. Due to the blood flow.
CT versus SPECT degree radius
CT does 180° SPECT does 360°
Two effects of spatial resolution
First the image is blurred the second effect is more subtle.
Scatter effects
The photons detected in SPECT and PET are electromagnetic radiation.
Two major types of interactions
Photoelectric effect and Compton scattering
The photoelectric effect results in…
Complete absorption of the photon, reducing the observed count rate, although scattered photons may still be detected.
To reduce Compton scatter
Use narrower pulse height windows
Ideal image peaks
Narrow tall peaks
Attenuation
Cold spot or loss of photons by a combination of photoelectric and Compton scatter
The higher the number of counts in a reconstructed image…
The less smoothing or cutoff a window function should have
One of the commercially available types of filter window combinations is the
Ramp Hann