Biophysical Techniques Flashcards

1
Q

Principles of diffraction

A
  • Arises from elastic scattering of radiation in directions other than incident direction
  • Diffracted waves interact with each other + produce interference patterns
  • Diffraction pattern from a single molecule has info needed to generate a model, but need large no for practical reasons e.g. signal to noise ratio
  • Crystal lattice = regularly repeated arrangement of molecules
  • Needed as scattering from individual molecules is too weak to measure
  • Crystal = 3D arrangement of atoms in real space w/ coordinates xyz
  • Its diffraction pattern = 3D array in reciprocal space w/ coordinate system hkl
  • The 2 are related by FT
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Diffraction

Bragg equation

A
  • In a crystal, only some angles give constructive interference + finite intensity, these spots = reflections
  • If x rays, incident at angle θ are scattered from adjacent crystal layers separated by a, there is a path-length difference btw scattering from adjacent layers that is 2p=2dsinθ
  • To get constructive interference btw 2 scattered waves, this path difference needs to be an integral no. of wavelengths. Gives Bragg equation: 2dsinθ = nλ
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Diffraction

Sources of radiation

A
  • Radiation must have a λ that is of the same order of magnitude as the crystal lattice
  • Produced in the lab by striking a target of metal w/ e-s accelerated at a high voltage
  • OR produced w/ synchotrons. High E e-s travel in a circular orbit in a vacuum
  • Orbiting e-s emit intense synchotron radiation at points of curvature
  • Crystal monochromator is used to select a wavelength appropriate for the experiment
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Diffraction

The sample

A
  • Usual samples = crystals or fibres
  • Crystallisation consists of 2 main steps:
    1. Nucleation = molecules dispersed in solvent start to gather into clusters
    2. Crystal growth = clusters need to reach a critical size before they become nucleus
  • Highly flexible + disordered regions are unfavourable
  • As the concentration of protein ↑ the protein becomes ↓ soluble until protein comes out of solution
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Diffraction

Vapour diffusion

A
  • 25ul of purified protein is mixed w/ equal amount of reservoir solution
  • Precipitant, like ammonium sulphate, binds H20 molecules + ↓ available H20, mimics ↑ protein conc
  • Solution is suspended on a droplet under a cover slip, which is put on top of reservoir
  • Vapour diffusion → net transfer of H20 from protein solution in the drop to the reservoir until precipitant conc. in both are equal
  • Drop shrinkage ↑ both the conc. of both protein + precipitant, moving conditions diagonally into nucleation
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Diffraction

Batch experiment

A
  • Crystal crops are incubated in low-density paraffin oil
  • Aqueous crystallisation drops are denser than the paraffin oil → stay beneath the oil where protected from evaporation + contamination
  • Protein is brought directly into the nucleation zone + samples are mixed at their final conc. at the start so conditions are constant + crystals x usually dissolve
  • x be used for crystallisation trials containing a small volatile organic molecules as they can dissolve into the oil e.g. phenol
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Diffraction

Other preparation for non-protein compounds in complex w/ crystal

A
  • Enables investigation of the structure of proteins interacting w/ ligands, like cofactors or inhibitors + heavy metal ions needed to solve the phase problem
    1. Co-crystallistaion: protein + crystal are crystallised together. Only method for producing crystals of protein in complex w/ large ligands like nucleic acids
    2. Soaking protein crystals in mother liquor that contains the ligand. Preferred when want to compare structure of pure protein w/ protein-ligand complex as crystals ↑ likely to be same form as pure protein
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Diffraction

Detectors

A
  • Area detectors are used
  • E.g. = charged coupled device (CCD)). Give fast readout times + are ideal for synchrotron use where higher throughput is required
  • Large phosphor screen is used to convert incident X-rays into optical photons which are transmitted directly to CCD w/ tapered optical fibres
  • The distance between two adjacent spots in a row/column is inversely proportional to the unit cell dimensions
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Diffraction

The phase problem

A
  • All detectors measure the intensity of the reflections
  • We can calculate the amplitude through its proportionality to the square root of the intensity
  • A wave is defined by amplitude, which we can infer, + phase, so there is a loss of information - the phase problem
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Diffraction

Isomorphous replacement

A
  • Patterson function = FT of I(hkl) {I(hkl) → p(uvw)
  • As I(hkl) = measurable quantity, P(uvw) can be obtained directly
  • Direct methods involving solutions by the Patterson function can be applied to small molecules + used to locate heavy atoms
  • For large molecules, P(uvw) gives v complicated patterns → need isomorphous replacement
  • Use a heavy metal ion (strong scatterer)
  • This changes intensities of the reflections from P(hkl) to Ph(hkl) P=protein, Ph = protein + heavy atom)
  • Changes in intensity can be measured + a Patterson difference map obtained from Ft of /p(hkl)-/ph(hkl)
  • If this map can be interpreted, we know phase + amplitude of Fh (hkl)
  • We can work out structure factors for Fh so we can work out amplitude of native protein /Fp/ + heavy atom derivative /Fph/ → solution for phase of each structure factor
  • Solving the structure like this gives 2 possible solutions for the phases, use at least 1 other heavy atom derivative
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Diffraction

Single + multiple isomorphous dispersion

A
  • Similar to IR but heavy atoms x have to be as large (often selenomethionine replaces methionine)
  • Takes advantage of capacity of heavy atom absorbing X-rays of specified lengths
  • As a result, Friedel’s law x hold so reflections hkl x = hkl
  • This inequality of symmetry-related reflections = anomalous scattering
  • In MAD, a complex data set is generated w/ native crystals giving /Fp/ for each of the native reflections
  • Data set using heavy atom derivative is collected at the same λ → /Fph/
  • 3rd data set is generated, maximises anomalous scattering by the heavy atom at a different x-ray length. Non-equiv Friedel pairs used to establish phases of reflections in heavy atom data + used to identify native phase
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Molecular replacement

A
  • Choice for determining the structure of a macromolecule for which is similar to another whose structure is already known
  • Calculation involves solution of the rotation + translation functions (rotated in 3D, calculate structure factors + agreement btw, then place in each position in unit cell)- Fact that split into 2 ↑ efficiency
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Diffraction

Refinement

A
  • Often, initial set of phases are determined + e- density map for the diffraction pattern is calculated
  • e- density map is used to identify portions of the structure that can be used to stimulate a new set of phases w/ higher precision
  • From this, a new e- density map can be created, more accurate atom positions can be derived → even better phase angle
  • Models can be improved by varying position x,y + z and mobility B to minimise residual by least squares method
  • New set of phases = a refinement
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Diffraction

Least squares refinement

A
  • An estimate of how good the agreement btw observed data (Fobs) + calculated (Fcalc) is given by
    R =Σ(Fobs-/Fcalc/)/ΣFobs
  • R of 0.2 is usually obtained for 2.5A resolution data. Lower res = lower R value
  • Oftem several cycles of refinement are needed
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Diffraction

Rigid body refinement

A
  • Proteins have domains w/ predictable structure, it’s possible to refine lengths of 2o structure, like a helices or B strands
  • All distances btw atoms are fixed so only 6 variables for each block of structure, somewhat simplified
  • Used for proteins that are difficult e.g. refinement of metabotropic glutamate receptor
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Diffraction

Validation

A
  • Typical Rfactor = 15-25%
  • Powerful check = Ramachandran plot (plot of the main-chain dihedral angles in a polypeptide chain)
  • Phi and psi angles are not used as restraints in refinement programs and hence their quality are independent checks of the model quality.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Diffraction
Temperature factors
Free R factor

A
  • Describes how much an atom oscillates/vibrates around the position specified in the model
  • All B factors higher than 50A^2 suggests observed atom is disordered

Free R-factor = important against incorrect model building
Here, 5-10% of reflections are quarantined from the refinement
Decreases = correct model building, increase = model is being built wrong
A decrease in the conventional R-factor at the expense of an increase in the free R-factor is diagnostic that the model is being overfitted,


How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

NMR

Overview

A
  • Detects reorientation of nuclear spin in an applied magnetic field, Bo
  • Advantages = ability to detect changes in chemical structure + environment and magnetic waves readily penetrate membranes + tissues w/o damaging
  • Only technique to report protein dynamics at atomic resolution
  • Disadvantage = fairly weak signals compared to other forms of spectroscopy due to small E separation btw E levels in nucleus
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

NMR

Background

A
  1. Spin
    - Nuclei in many isotopes have spin e.g. 13C
    - Nucleus w/ spin has a magnetic moment, un, that can interact w/ an applied static magnetic field. Bo
    - If nucleus has spin I, interaction w/ Bo means there are 2I + 1 possible E levels
  2. Bo field
    - In a typical NMR experiment, a strong Bo field is applied to the sample
    - This Bo is generated by a current in a solenoidal coil made of ‘superconductor’ wire
  3. B1 field
    - In addition to static Bo, rotating B1 is needed to induce phase coherence in the x-y plane
    - Usually generated by a resonator coil
    - Coil is tuned to resonant frequency of the signal to be detected
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

NMR

Spectral parameters

A
  • An NMR spectrum is characterised by a series of lines or resonances
  • There are 5 parameters that define such lines
  1. Intensity
    - Absolute conc. x be obtained reliably, but relative conc. can be obtained accurately from measurement of resonance intensity
  2. Chemical shift
    - Nuclei are surrounded by e-s; Bo induces currents in the e- clouds that ↓ effective field experienced at the nucleus
    - Reference compound, usually tetramethylsilane
    - Intrinsic shift = characteristic of a particular 2o structure,
    - Induced shift = arises from influence of environmental effects. Local environment shifts allow resonances of groups w/ same chemistry to be resolved
    - References for a-helices and B sheets have also been assembled
    - Forms basis of chemical shift index which compared to CD spectra = info at residue level
  3. Spin-spin coupling + multiplet structure
    - Spin-spin coupling = magnetic interactions btw neighbouring, non-equivalent NMR-active nuclei
    - Measured by spin-spin coupling constant J
    - 3 equivalent nuclei give rise to a quartet w/ intensity ratio 1:3:3:1, 2 equivalent = 1:2:1
  4. T2 relaxation time
    - T2 = time constant of the decay of Mxy components
    - Line width of an NMR signal is determined by T2 - short T2 = broader lines
    - T2 relaxation = caused by transient magnetic fields at any frequency. So, T2 ↓ as molecular reorientation rates slow
  5. T1 relaxation time
    - T1 = time constant that describes recovery of M2 (magnetisation along Bo field direction) after an applied perturbation
    - T2 < T1 as return magnetisation to z direction inherently causes loss of magnetisation in xy plane
    - Short T1 = magnetisation recovers rapidly + a spectrum can be acquired in ↓ time
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

NMR

Nuclear overhauser effect

A
  • A dominant relaxation mechanism = dipolar interactions with other spin magnetic moments
  • An important consequence of relaxation = nuclear overhauser effect, which can be used to determine intra-molecular distances
  • NOE effect is the change in population of one nucleus when another magnetic nucleus close in space is perturbed by irradiation
  • Short range through-space interaction
  • Can divide inter-atomic distances which can help confirm the 3D structure of a molecule
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Types of NMR

1D

A
  • Has 2 dimensions: X axis corresponds to the frequency axis (chemical shift in ppm) + Y axis = intensity
  • From 1D NMR, can work out folded state easily from TMS region - if unfolded, no structured hydrophobic region
    Can tell how close to a e- -ve atom through shift
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Types of NMR

2D

A
  • Has 2 frequency axis
  • Building blocks of preparation, evolution, mixing and detection
  • Involves a series of 1D experiments in which a single decay has been altered in length
  • Both evolution + detection are time periods, t1 + 2, during which chemical shift + scalar couplings evolve

[1H-1H] COSY Experiment

  • Through-bond NMR
  • Correlates nuclei via their scalar couplings
  • Consists of a single RF pulse → specific evolution time → 2nd pulse → measurement period (t2)
  • Protons that are more than 3 chemical bonds apart give no cross signal because the 14J coupling constant are close to 0
  • Cross signal btw H^n + H^a are important - this phi torsion angle of the protein backbone can be derived from the 3J coupling constant btw them
  • There are 2 types of peak:
    1. diagonal peak (have same frequency coordinate on each axis + appear along the diagonal)
    2. Cross peak (have different values for each frequency + indicate couplings btw pairs of nuclei

[1H-1H]-NOESY experiment

  • Examples of through-space experiment
  • Uses dipolar interaction of spins (NOE) for correlation of protons
  • The intensity of NOE is proportional to 1/r^6
  • Normally a signal is only observed if distance is smaller than 3A
  • Correlates protons that are distant in aa sequence but close in space

Major advantage of 2D spectra = improved spectral resolution + easier identification of correlation btw peaks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Types of NMR

3D

A
  • Can easily be constructed from a 2D experiment by adding additional evolution time + 2nd mixing period btw 1st mixing period + direct data collection

Triple resonance

  • Used for assignment of larger proteins
  • Triple resonance as 3 different nuclei (1H,13C,15N) are correlated
  • Only contain a few signals on each frequency, so issue with spectral overlap ↓

2nd type involves using 2 tandem NMR experiments e.g. 1H-NOESY-TCOSY spectrum where t1 is NOESY evolution time and t2 = TCOSY mixing time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

NMR

Refinement

A
  • Calculation of structures is done by incorporating various observed NMR restrains into a molecular dynamics simulation protocol
  • MD simulations calculate the positions + velocity of protein atoms in a series of small steps
  • Several 100 structures are produced + structures of a subset that has the lowest E are superimposed
  • Following computation of structures, likely a no. of experimental ambiguities + mis-assignments so cycle of iterative corrections
    Many visualisation software packages that perform chemical shift assignments automatically e.g. SPARKY 3
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

NMR

Size limit + solutions

A
  • In order to detect through-bond correlations, need good linewidth compared to values of coupling constant
  • In 1H spectra of small proteins, J is large enough for through bond correlations to be detected for atoms up to 3 bonds apart
  • As molecules ↑ in Mw, they tumble more slowly in solution → faster T2 relaxation → ↑ linewidth
  • Observations in through-bond experiments e.g. COSY are more difficult when linewidth is similar to coupling constant
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

Solution to NMR issues

A
  • Extend into 3D: only a few signals are contained, ↓ the issue of spectral overlap
  • Deuteration: swap 1H for 2H. Simplifies the spectra + gives better relaxation properties due to the fact that 2H nucleus has a much smaller magnetic dipole than 1H
  • TROSY: useful pulse sequence that can extend the accessible molecular weight range. When 1 doublet is broader than another due to chemical shift anisotropy, TROSY can be used to detect the narrower signal. Works best in high magnetic fields
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
28
Q

MRI

A
  • An image of water protons in the specimen is obtained by collecting spectra while different field gradients are applied to the sample
  • These field gradients ‘label’ the position of sample components in space, which allows images to be constructed
  • Contrast MRI
  • To generate a good image, it’s important to have intensity variation in signals from different parts of a sample MR I
  • One way to get contrast = exploit different T1 + T2 values in a sample
  • Functional MRI = 2 MRI images are acquired in quick succession whilst the subject forms specified brain tasks
  • Difference signals are superimposed
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
29
Q

Mass spectrometry

Ionisation

A
  • 2 most common techniques = MALDI + ESI
  • Complementary techniques: both can analyse molecules with mass > 1mDa
  • MALDI ionises samples in solid state, ESI in liquid
  • MALDI is ↑ sensitive + more salt tolerant than ESI, but ESI is good for detecting native states + different conformations
  1. MALDI
    - Ionisation is initiated by irradiation of the sample with a laser beam
    - Solvent evaporates leaving matrix w/ analyte spread throughout the crystals + an applied laser pulse vaporises the matrix containing the analyte, creating ions which can be passed to mass analyser
  2. ESI
    - Liquid containing the macromolecule of interest is sprayed through a capillary to create a fine aerosol
    - For large molecules, ESI mainly forms multiply-charged species
    - As mass spectrometers measure m/z ratio, production of species w/ high z value is advantageous as molecules w/ ↑ m can be measured w/ instruments w/ ↓ mass range
30
Q

Mass spectrometry

Analysis

A
  • Once ions are formed by MALDI/ES, they pass to a mass analyser where they’re separated on the bases of m/z ratio
  1. Time of flight
    - Ions with different mass have different transit times
    - After laser pulse + extraction of ions by an electric field, E1, a field E2 accelerates the ions into the drift region
    - Using the equation for the kinetic energy of the ion, the length of the tube and transit time, it becomes apparent the transit time is proportional to (m/z)^12
  2. Quadropole Analyser
    - Consists of 4 parallel cylindrical rods
    - Filters ions on the basis of their m/z ratio using oscillating electric fields applied to the rods
    - Ions w/ certain m/z values only reach the detector at particular values of applied voltage
  3. Quadropole traps
    - Powerful alternative to analysers that are on a moving ion beam (above) = trap ions selectively
    - Electric fields are applied in 3D. The shape of the 3 electrodes is designed to trap ions in space btw them when certain electric fields are applied
    - Voltage is applied + trapped ions oscillate w/ characteristic frequencies. Applied voltages tuned to these frequencies cause certain ions to have enough kinetic E + are ejected in a mass selective manner
  4. Orbitrap
    - Ions are injected into a shaped electrode system w/ an electric potential applied btw inner + outer electrode surface
    - Ions become trapped as their electrostatic attraction to the inner electrode is balanced by centrifugal force
    - By sensing ion oscillation, the trap is used as a mass analyser
  5. Tandem MS/MS
    - 1st stage of separation is achieved when 1 mass analyser (MSI), followed by further analysis of the separated ions in a 2nd mass analyser (MS2)
  6. Fragmentation
    - Collision chamber containing a natural gas e.g. argon is used to produce peptide fragments
    - When an ion collides w/ a neutral gas, enough E can be transferred to cause chemical bonds to break. This is collision-induced dissociation. Mainly CO-NH bond breakage
    - If a double charged peptide ion is fragmented, both b+y ions are formed. When singly charged ions fragment, either a,b or y is formed
31
Q

Mass spectrometry

Detectors

A
  • Ion detectors, usually some kind of photon or e- multiplier, records either charge induced or current produced when an ion passes by or hits a surface
  • Mass accuracy = typically better than 0.01%
32
Q

Mass spectrometry

Data analysis

A

In MALDI, ionisation occurs by addition of 1H+ to the molecule so 1 dominant peak at full mass w/ 2o at m/z

  • ESI = more complicated as more ions w/ different charges are produced
  • Addition of one positive charge corresponds to addition of 1H+ (m/z = H +nH+)/n
33
Q

Isothermal titration calorimetry (ITC)

Instrumentation

A
  • 2 adjacent cells, a reference + sample cell, surrounded by a thermal insulation jacket
  • Reference cell is kept at constant temp by feedback control to the jacket, keeps temp difference btw reference cell + jacket constant
  • High concentrations of protein are needed
  • ↑ amounts of ligand are added at fixed time intervals
  • Any temp change btw sample + reference cell is sensed + power is fed in to keep temp change at 0. Data fed to computer
  • As the course of infection continues, the sample binding sites gradually saturate + the exothermic effect disappears
34
Q

ITC

Binding measurements

A
  • 1 spike = 1 injection
  • For the 1st few injections, small molecule is less abundant so immediately all bound by protein, releasing max heat
  • Early spikes = deltaH
  • Slope in middle of sigmoidal curve= Ka
  • Provided the conc. of L + P are known, ΔH, Ka + n (no. of C that bind P) can be determined by curve-fitting to the ITC data using a suitable model
  • Need a significant, measurable value of ΔHobv + that there is suitable curvature of ΔHobvs vs molar ratio plot
35
Q

ITC

Kinetics

A
  • Universal measuring method, x require chromophore or labelling + opaque solutions x a barrier
  • Dependence of rate on substrate conc. = v = (Kcat[S])/(Km + S)
  • ITC measures the thermal power generated by enzymatic conversion of substrate to product, which can be measured experimentally w/ ITC
  • Michaelis M curve can be generated from 1 ITC experiment as a wide range of [S] values can be sampled during the experiment
36
Q

Surface plasmon resonance

Overview

A
  • When incident light moves from medium of relatively ↑ refractive index to ↓ one, the ray of light reflects
  • When the light is entirely reflected, TIR occurs → evanescent wave in the medium that has the lower refractive index which decays with distance from the interface
  • In SPR, the evanescent wave excites electrons within the metal layer of a metal–dielectric interface, yielding surface plasmons.
  • Surface plasmons = electromagnetic surface waves that propagate parallel to the interface region.
  • As the plasmon waves penetrate into the medium with the lower refractive index, shifts in the intensity of the reflected “angle” of polarised light is recorded.
  • The reflected light intensity is calculated as a function of the incident light angle A shift in the SPR angle by 0.0001 degree corresponds to one unit shift in SPR signal
  • Charge coupled devices = detection system
  • Ligand is immobilised to surface of SPR transducer.
  • When molecule binds ligand, refractive index changes at interface + SPR takes place at different angle
  • Advantage = - x need a labelled molecule + associate + dissociation can be measured in real time
37
Q

Surface plasmon resonance

Experimentation

A
  • Kretschmann configuration = light source, prism, gold film + detector
  • Medium 1 (higher refractive index) = prism, medium (2) = solution of interest
  • Ligand is immobilised on the sensor surface whilst analyte is free in solution + passed over the sensor. V small amounts of sample are required
  • The SPR signal is related to mass changes at the sensor chip surface.
  • SPR devices = non-selective, control experiments are important
  • Immobilisation of ligand to the surface = by covalent coupling (can be used for any bio molecule) or indirect by a covalently coupled molecule (only used w/ molecules w/ a suitable binding site)
38
Q

Surface plasmon resonance

Kinetics

A
  • Sensogram plot of response vs time
  • Shows interaction btw 2 molecules: if interaction occurs, refractive index at the surface changes + there is an ↑ in signal intensity, 1RU = 10^-4 degrees
  • Have association phase. steady state + dissociation phase
  • Kd can be determined from the ratio of association + dissociation rates that can be observed directly
  • SPR response correlates with a change in mass conc. on sensor surface + depends on Mw
39
Q

Fluorescent polarisation

A
  • Measurements of FP give info about motion + molecular interactions
  • If fluorophores are excited w/ plane polarised light + the fluorescence is observed through analysing polarisers, light emitted = largely depolarised
  • Due to re-orient. of fluorophore during lifetime of its excited state caused by rapid brownian motion
  • Due to photo selection: only flurophores w/ components of their transition dipole moments oriented along the direction of polarisation are excited
  • Fluorescence anisotropy = phenomenon where light emitted by a fluorophore has unequal intensities along different axis of polarisation
  • If molecules rotate during the time btw absorption + emission, observed A becomes less than Ao so motion causes depolarisation
  • The reduction in polarisation depends on the degree of motion: free labelled ligand has fast tumbling so fast depolarisation (low A), bound ligand ↑ in mass so ↓ speed, ↓ depolarisation, ↑ polarisation
40
Q

Analytical Ultracentrifugation

Overview

A
  • Sedimentation force is generated by placing a sample in a spinning rotor
    Application of centrifugal forces causes the depletion of a macromolecule at the meniscus and the formation of a concentration boundary that moves towards the bottom of the centrifuge cell as a function of time.
    Diffusion causes the sedimenting boundary to spread with time, monitor this motion + shape of a boundary, Use optical system
41
Q

Analytical Ultracentrifugation

Results

A
  1. Sedimentation velocity
    - Measurements of sedimentation properties of molecules in solution tell us about mass, shape, interaction w/ themselves + others
    - Spinning rotor → sedimentation force of mw^2r
    - At the ‘terminal’ velocity, sedimentation force is balanced by frictional force + the mean velocity is constant
  2. Sedimentation coefficient
    - Measured in Svedberg (S)
    - S can be measured by viewing the sedimenting molecules at ↑ time intervals
    - Unfolded/highly elongated shapes = more hydrodynamic friction so ↓ sed coefficient than folded
    - As sample conc ↑, generally observed S ↓ due to interference from other molecules
    - If there is an association, higher S are observed at ↑ conc. as amount of higher mass complex ↑ w/ ↑ conc.
  3. Shape + f/fo
    - Functional coefficient, f, of an object moving through solution is related to its shape
    - If object volume other than a sphere, f/fo is greater than 1
    - Ranges from 1.3 for globular hydrated proteins to 2.0 for elongated/glycosylated proteins
    - Radius of gyration (from MALS) can be compared w/ hydrodynamic radius from DLS or AUC ⭢ shape factor P
    P = 0.778 for hard sphere, 2.36 for stiff rod
  4. Polydispersity
    - Sedimentation velocity can be used to analyse mixtures of species + detect interactions btw
    - Plot c(s), amount of material in a sample, w/ a sedimentation coefficient btw S + δs gives info about distribution of observed sedimentation coefficients
  5. Ligand binding + association
    - Faster-migrating complexes remain in a pool of slower sedimenting species so associations + dissociates can take place through time course of the experiment
    - Data set from a single sedimentation run, can deduce a curve + calculate Kd
42
Q

Analytical Ultracentrifugation

Sedimentation Equilibrium

A
  • At lower motor speeds, sedimentation of molecules down the centrifuge cell is balanced by their diffusion back up the cell
  • Can be long, helped w/ shorter column can be used
  • Plot of c vs r^2 gives a straight line with slope = M
  • Used to detect complexes
43
Q

Dynamic Light scattering

Overview

A
  • Particles or macromolecules suspended in a liquid medium undergo Brownian motion which causes fluctuations in local conc.
  • Results in local variations in refractive index + fluctuations in the intensity of the scattered light
  • If a laser source is used as the irradiation light source, these time-dependent fluctuations can be analysed
  • Large particles diffuse slower than small ones so scattering fluctuations have fewer high freq. components
44
Q

DLS

Analysis

A
  • Scattering vector q is used to determine D, which allows to calculate Rh from Stokes-Einstein equation
  • Can also measure frictional ratio (how much shape deviates from a sphere), as deviates f/fo ↑
  • Molecular weight can also be approximated using Rh, specific particle density, hydration and Avogadro constant
45
Q

Circular dichroism

Overview

A
  • Difference in absorption of left vs right circularly polarised light
  • CD = for optically active materials, but can also be induced via covalent bonding to a chiral chromophore
  • Chromophore needs to be intrinsically asymmetric, or placed in an asymmetric environment to produce a CD signal.
  • Therefore LH CPL propagate through a chiral sample at a different speed to the RH counter part
46
Q

Circular dichroism

Analysis

A
  • Proteins present CD bands in two different spectral regions, the far-UV or amide region (175–250nm) and the near-UV or aromatic region (250–300nm)
    Far-UV CD informs about the secondary structure content. E.G. = alpha helix structure displays the most invariable band pattern: a characteristic spectrum with a positive band at 190nm and two negative bands at 208 and 222nm
    Near-UV CD reflects changes in the environment of aromatic side chains.
  • Requires curve fitting procedures based on a set of reference spectra w/ know 2o structure components
  • a helix, B sheet + coiled coil all have characteristic shape
  • LOW res/ idea of structure e.g. poly-lysin = 50% a helix
47
Q

Multiangle light scattering

A
  • Single frequency, polarised light beam is used to illuminate a solution w/ macromolecule of interest
  • Multiple detectors placed around a flow cell at several fixed angles
    For larger macromolecules, more and more light is scattered in the forward direction at smaller angles. Need multiple angles
  • When there are many molecules in solution, each macromolecule scatters light via the induced dipole mechanism so intensity of scattered light is proportional to conc. of macromolecules
  • More polarisable molecules = larger induced dipole = ↑ intensity of scattered light
  • SEC-MALS uses size-exclusion chromatography coupled with an in-line MALS detector + an in-line UV/Vis detector
48
Q

MALS

Analysis

A
  • For larger macromolecules, MALS can be used to measure Rg
  • Rg found by plotting 1/Rθ vs sin^2(θ/2)
  • SEC-Mals can be used to determine molecular mass of proteins + complexes
49
Q

SAXS/SANS

Overview

A
  • Small angle x-ray / small angle neutron scattering
  • Solution of macromolecules = exposed to X-rays (λ = 0.15nm) or neutrons (λ=0.5nm)
  • Scattered X-rays are detected by an area detector + observed angular dependence of the scattering is translated into plot of I vs Q (tells about size + shape)
  • SANS: at certain ratios of H20:D20 (match points), scattering from the molecule = that of the solvent (so is eliminated when scattering from the buffer is subtracted e.g. 43% D20 for protein
  • When a particle is made from ↑ material e.g. ribosome or when SLD of continuous phase is same as SLD of object, various parts can be studied selectively by changing D20/H20
  • Disadvantages = info from SAS is limited by spherical averaging so reconstructing 3D structural models is inherently ambiguous. Also, SAXS = low-resolution method + doesn’t provide info at an atomic level. Resolution - around 20A
50
Q

SAXS/SANS

Analysis

A
  • Rg can be determined from the angular dependence of scattering at small angles using Guinier plot
  • Plot Q vs lnI
  • Lower Q region = characteristic of dimension of particle
  • Make Guinier plot of lnI vs Q^2
  • Molecular weight is proportional to I at X intercept
  • Radius of gyration is proportional to slope (slope = -Rg/3)
  • Rg (sphere) = sqrt(3/5R) rod = sqrt(1/12L)
51
Q

Optical microscope

A
  • Abbe limit = 0.61λ/NA. Describes resolving power of a circular aperture
  • Light passes through sample in a transmission microscope
  • Lenses are imperfect + usually have aberrations associated w/ them e.g. chromatic aberrations which results from different wavelengths being refracted through the lenses differently
  • Staining highlights features e.g. cosin = acidic dye that stains cytoplasm pink
  • Brightfield microscope = example that illuminates a sample by white light from below
  • Resolution limited to ~200-300 nm due to diffraction of light on the aperture of the microscope
52
Q

Fluorescence microscope

Overview

A
  • Excellent contrast as light is detected from specific fluorophores at a wavelength that is different to source of illumination
  • For fluorescence, common DNA stain = DAPI w/ blue emission maximum
  • Fluorescent proteins like GFP can be attached to proteins
53
Q

Fluorescence microscope

FRAP, FLIP + FRET

A
  • FRAP (fluorescence recovery after photobleaching( = recovery of fluorescence in a defined region of a sample after photobleaching
  • Selected area of the cell is briefly illuminated with high intensity laser light + subsequent recovery of fluorescence is monitored as a function of time + is from movement of unbalanced fluorophores from surrounding area into bleached region
  • Regeneration time is a measure of fluorophore diffusion rate
  • FLIP (fluorescence loss in photobleaching)
  • Complimentary technique to FRAP
  • Measures ↓ of fluorescence in a defined region adjacent to a bleached region
  • Progress of fluorescence decay in area adjacent to bleached area is monitored as a function of time
54
Q

Fluorescence microscopy

Confocal microscopy

A
  • Laser beam is continually scanned across and down the sample and the image is constructed from discrete pixels that are compiled as the laser scan proceeds.
  • Eliminates out-of-focus info that can sometimes occur w/ a ‘pinhole’ in front of image plant
  • As only 1 point in the sample is illuminated at a time, 2S or 3D imaging requires scanning over specimen
  • Spinning disc confocal uses multiple excitation + emission pinhole apertures on a rapidly spinning disc. Pinhole array can sweep entire field of view over 1000 times per second
  • High speed → 360 frames per second + can ↓ peak excitation which in turn ↓ photobleaching
55
Q

Fluorescence microscopy

TIRF

A
  • Illumination that is totally internally reflected generates an evanescent wave that rapidly decays in the sample
  • Evanescent wave can be used to excited a few fluorescent molecules near the surface of the coverslip
  • Good for visualising events that take place near the surface of the cell e.g. cell adhesion
56
Q

Fluorescence microscopy

Super resolution microscopy

A
  • Spot produced in confocal has a 3D point spread function with a finite diffraction-limited size
  • When scanned, only have resolution slightly better than conventional microscope limit
  • Super resolution techniques allow images to be taken w/ a resolution higher than imposed by diffraction limit
  1. Localisation based technique
    - STORM/PALM relies on cycles of activation, excitation, photobleaching + localisation of a small no of labelled molecules that are separated through a distance greater than Rayleigh limit
    - PALM = specific fluorophores are activated by lasers + emit for a short time, but eventually bleach
    - Laser stochastically activates fluorophores until all have emitted
    - Multiple snapshots are taken on an EMCCD camera
    - STORM = red laser used to turn all fluorophores ‘off’, green laser turns a small + random no. ‘on’
    - After a set no. of repetitions, individual images are processed + merged into a final image
    - Allow an increase of optical resolution by a factor of 10 up to ~20 nm in lateral and ~50 nm in axial direction.
  2. Patterned illumination techniques
    - Structured illumination microscopy (SIM) = fluorescent sample is excited multiple times using striped illumination patters, changing position + orientation of stripes each time
    - Stripes fired at the sample interact w/ high frequency light produced from the sample → 3rd pattern which can easily be analysed
    - Resulting patter analysed w/ software based on Moire pattern, removes excitation pattern
    - STED
    - 30nm resolution, sample irradiated w/ a normal excitation spot + ring of light at a wavelength that stimulates depopulation not some of the excited fluorophores
    - Applied at an effected wavelength so light ring essentially quinces fluorescence from the area so the combined irradiation from the ring + objective lens gives smaller spot
    - High damage due to intense laser. Improves x+y not z
57
Q

Electron microscopy

Transmission electron microscopy

A
  • Electrons are produced with emission ‘guns’ + are attracted to a +ve anode w/ ↑ voltage
  • Electrostatic and/or electromagnetic lenses are then used to focus the e- beam to form an image
  • Radiation damage in EMs is a serious problem as exposure to an e- beam causes significant damage to delicate biological structures
  • Detectors can be CCDs which are good for lower resolution images
  • NA values can be 0.009 giving resolution of 0.26nm

Sample preparation

  • Harsh treatment requires good preservation, chemical fixation + dehydration with organic solvents
  • Staining is used as intrinsic contrast in the EM is low
  • Most common - metals like tungsten. If sample binds metal = +vely stained. More commonly if binds background = -vely stained

Cryofixation

  • Sample rapidly cooled to 77K
  • Sample preserved in a snapshot of its solution state by cooling so rapidly the water forms a vitreous solid
  • Solution containing the particles of interest is suspended as a thin layer across a hole in the EM grid + the grid is held in tweezers mounted on a plunger that falls into liquid at 77K
  • Observe samples in close to native state as x fixing or staining

Tomography

  • Specimen is tilted at different angles w/ respect to the e- beam + a series of different projections made
  • Tilt usually restricted to +-60
  • Means after reconstruction, resolution in different directions x the same as data are missing in some regions

Image processing

  • Used to improve noise (noise arises from contributions to the image that are x directly from specimen)
  • Background noise = v high as low e- dose needed to avoid damage to sample
  • Reconstruction = important
58
Q

Electron microscopy

Molecular e- microscopy

A
  • To get ↑ resolution, x staining, low e- dose + v low temps are used to ↓ e- beam damage
  • Allows observations to be made at the molecular level
  • E.g. single particle studies
  • Determines the structure of macromolecules from images of individual particles
  • Crystals x needed, v small amounts of material used
  • Resolution depends on sample but can be 1nm
  • 3D structure is made from noisy 2D images by processing with a computer
59
Q

Electron microscopy

Scanning electron microscope

A
  • Fine beam of e-s is scanned back + forth across the sample
  • Gives image 3D effect
  • Unlike TEM, magnification is x dominated by lens
  • Condenser + objective lens focuses the beam to a spot rather than image the specimen
  • Living cells + tissue usually need chemical fixation to reverse
  • Another advantage = analytical techniques can be applied like x-rays to identify chemical composition
  • Resolution = 3mm
60
Q

Optical tweezers

A
  • Uses light to manipulate microscopic objects
  • Radiation pressure from a focused laser beam can trap small particles
  • Laser beam is focused by a high quality microscope objective to a spot in the specimen plane
  • This creates an ‘optical trap’ that can hold a small particle at centre
61
Q

Optical tweezer

Theory

A
  • Light creates a momentum that is prop. to its E in direction of propagation
  • Conservation of momentum requires the object undergo an equal + opposite change → force on the object
  • When this light interacts w/ a bead, light rays are bent
  • 2 forces result:
    1. If scattering force of refracted rays > reflected rays, a restoring force is also made in a plane → stable trap
    2. Gradient force is a restoring force that pulls bead to centre xy plane. Arises from gradient of Gaussian intensity profile
  • For stable trapping in all three dimensions, the axial gradient component of the force pulling the particle towards the focal region must exceed the scattering component of the force pushing it away from that region.
  • So, need a steep gradient in the light, produced by sharply focusing the trapping laserbeam to a diffraction-limited spot using a high NA
62
Q

UV/Vis spectroscopy

A
  • When white light passes through an object, some wavelengths are absorbed
  • When sample molecules are exposed to light w/ E that matches a possible electronic transition within the molecule, some of the light energy will be absorbed as the electron is promoted to a higher energy orbital.
  • An optical spectrometer records the wavelengths at which absorption occurs
  • The resulting spectrum is presented as a graph of absorbance (A) versus wavelength,
  • In inorganic molecules, only certain functional groups that have valence e-s of low excitation E can absorb UV + visible radiation
  • The more easily excited e-s, longer λ can be absorbed
  • In addition to organic compounds, transition metals can also be analysed
  • Can use beer lambert law A = ecl
63
Q

IR spectroscopy

A
  • Used to study + identify chemical substances + functional groups
  • Exploits molecules absorb frequencies that are characteristic of their structure
  • These absorption’s occur at resonant frequencies- frequency absorbed radiation matches vibration frequency
  • The IR spectrum is recorded by passing a beam of IR light through the sample
  • When the frequency of the IR is the same as the vibrational frequency of a bond = absorption
  • Triple bonds have higher stretching frequencies that double bonds, which have higher frequencies than single bonds
64
Q

IR spectroscopy

Sample preparation

A
  • Gas: need long path length sample cell. Sample gas concentration below ppm can be measured w/ a White’s cell- IR light is guided w/ mirrors to travel through glass
  • Liquid: sandwiched btw 2 transparent plates of salt
  • Solid: many ways e.g. crush sample w/ mulling agent e.g. oil nujol
65
Q

IR spectroscopy

FTIR

A
  • IR guided through an interferometer then through sample
  • A moving mirror changes the distribution of IR that passes through interferometer
  • Signal directly recorder
  • Ft turns raw data to light output as a function of IR wavelength
66
Q

Diffraction

Electron density map resolution

A

Electron density map
Resolution
As the scattering angle increases, the separation of reflecting planes decreases so scattering objects close in space can be resolved
At low resolution (8-3.5A), the overall shape of the molecule can eb seen
At medium res (3.5-3.5A) aa sc can be identified
At high res (2.5-1A) individual atoms can be located

67
Q

Mass spectrometry

Proteomics

A

Identification
Protein spot on 2D gel can be excised, enzymatically degraded (e.g. w/ trypsin), separate fragments w/ liquid chromatography + analyse w/ MS
If identification not possible at this stage, peptides subject to tandem mass spec + fragmentation to give sequence info
Quantification
Can identify changes in protein levels btw 2 samples
1. Label-free approach = 2 proteomes are analysed separately w/ LC-MS/MS experiments. Peptides are identified in different LC runs from their retention times + m/z values
2. Isotope labelling approaches = ‘heavy’ stable isotopes are incorporated into the sample and ‘light’ into the control. Isotopes can be introduced at different stages of preparation e.g. 13C6Arg and 12C6 Arg, as Arg has 6C, peptides from sample w/ heavy isotope = 6Da heavier than Arg in control proteins.
Means can quantify differential protein expression in different samples

68
Q

DLS

Applications

A

Determine size distributions of a wide size range
Can determine solution viscosity applying a sample with know nRh + shifts to higher or lower rH proportional to the viscosity difference according to Stokes-Einstein eq.
Macromolecule degradation + disassembly
Homo and hetero-oligmerisation
E.G. = pH dependent dissociation and hydrolysis of Apoferritin

69
Q

Super resolution Microscopy

Atomic Force Microscopy

A
  • Uses a cantilever w/ v sharp tip to scan over a sample surface
  • As the tip approaches the surface, the close-range, attractive force btw the surface and the tip causes the cantilever to deflect to the surface
  • Laser beam detects cantilever deflection to or away from the surface
  • Images the tomography of a sample surface by scanning the cantilever over a region of interest
  • Lateral resolution of AFM is low (~30nm) due to the convolution, the vertical resolution can be up to 0.1n
70
Q

Optical Tweezers

Application

A

Applications
Capture, separation
+ assembly of cells

+ By capturing tiny particles e.g. biological cells, viruses, Brownian motion of the particles can be overcome
+ fixed in the field of the microscope for the researcher to observe

+ Stable captured particles can be moved to a specific pattern + arranged in a regular pattern. -can assemble particles
+ cell arrays

+ Can measure interaction btw particles/cells by measuring mechanical properties
Biomolecules 

+ Basic laws of life movement are explained by measuring the physical forces such as the tiny force of biological single molecule
+ motion step size

+ In order to manipulate a single molecule, need to connect the molecule to the microsphere
+ indirectly manipulate 

+ E.G. the 2 ends of the DNA chains are connected to two microspheres, and the microspheres are manipulated by a double-beam tweezers to stretch the DNA molecular chain and measure its elastic properties
-Optical rotator

+ Not only captures the Microparticles but allows angular rotation of micropraticle 

+ In order to achieve the rotation of the particles, the optical rotator requires a special beam of angular momentum, such as a Laguerre-Gauss beam

+ Rotating live cells can be imaged at various angles, observe full 3D appearance