Audiometric Testing and program equipment Flashcards
Equipment for an audiometric test
- audiometer
- sound attenuating booth
- sound level meter with the ability to measure in octave bands
- calibration equipment
Audiometer
- instruments used to measure hearing threshold
- electro acoustical instruments that provide pure tones at several frequencies and a range odd sound levels
- 500, 1000, 2000, 3000, 4000, 6000, 8000 HZ
- times are transmitted to the subject via the headphones which are calibrated to that specific audiometer and cannot be used with another machine
- 3 types
Manual audiometer
- tone presentation, frequency, and sound level, and recording of the threshold is done by the technician
- models are reliable and easy to operate
Manual audiometer advantages
- least expensive
- technician has control over the tone operations
- can manually fix if a person has trouble responding to the tone frequencies
Manual audiometer disadvantage
- if the examiner fails to use appropriate standardized technique human error can obscure the results
- testers can develop a rhythm that is easy to predict
- testing large number of workers can be tedious
- distracted humans can easily mess up
Microprocessor audiometers
- most common
- microchip is programmed for tone presentation, frequency selection, and sound level selection
- records responses based on hearing thresholds
- tone is played and worker presses a button
- if responses are inconsistent the machine will indicate the thresholds that need retesting
Microprocessor audiometer advantages and disadvantages
- can also be operated manually
- can download audiometric data of workers tests onto computers
- times are consistent and eliminates technician error
- prints results
- expensive
Computer controlled audiometer
- not self contained
- connected to a computer through a linking device
- software is loaded onto the computer to do the test
- the technician should not feel comfortable because the machine does all the work , results aren’t automatically valid
Difficulties testing with audiometer
- patients could have tinnitus and blend into the tones so the technician needs to override the system
- machine could stop working and technition needs to enter tones manually
Considerations when choosing an audiometer
- interface to a computer
- manual option
- storage
- price range
- portability
- printer capability
- fault cancelation and function checks
Storing and handing audiometer
- stored in a dry cool place at room temp
- covered and protected from dust
- earphones stored with cushions facing each other
- cleaning with soap and water
- no fluid attend electrical components
- earphones not picked up by the cord
CSA standards for a fixed facility
- a permanent testing area away from busy hallways and noisy equipment
CSA standards for a mobile facility
- built into self-contained portable facilities that can move to a temporary location
What does a testing area need to include
- a testing area with a chair , earphones, response button, required noise monitoring equipment
- a control area with an area for an audiometer and technician which includes a chair, desk, mirror , divider in which the person being tested cannot see the technician
- an advising area to discuss results and post testing procedures
Testing difficulties in the environment
- when there’s lots of people to be tested, each should be isolated by separate testing booths or partitions
- can be heard in noisy environments because workers can be distracted and have difficulties concentrating
Background noise in testing
- can interfere with testing and cause hearing to appear worse than it is
- must be measured before you move to a different location and records and calibration must be kept for that location
Sound analysis
- is done to ensure acceptable noise levels in certain testing facilities
- can be measured using an octave band analyzer and a sound level meter
- a “ regular” sound level meter measures sound in decibels and produces one reading
- an octave band analyzer can separate the sound into different frequencies and measures the decibels at each frequency
- also a separate dial for changing the settings for each frequency so that the individual decibels reading can be made for each frequency that contributes to the overall sound
Sound attenuating booths
- ensures a quiet testing environment
- make sure the sound booth is appropriate for the environment
Supra-Aural earphones
- have an earphone cushion and headband
- noise reducing headphones can quiet the area
- must be calibrated to the audiometer
Maintenance and Calibration
- must be calibrated using the CSA standards
- daily check and laboratory calibration
- checks help to ensure the validity of testing
Functional daily calibration check
- visual check to makes sure wires, plugs, leads, and earphones du not have any damage
- ensure an dials, buttons, knobs, and lights work
- connections
-warm-up the machine and put on earphones
Frequency listening check for daily calibration check
- set the audiometer at 60 dB and check tones at all frequencies in both earphones
- ensure no clicks
- make sure there is no cross talk
- make sure signal is not broken
Intensity listening check for daily calibration
- set frequency at 4000Hz
- set audiometer to right ear
- present tones at each decibel level starting at 60db and ending at 0
- change to left ear and repeat step 3
- listen to the volume increases and decreases properly with no signs of noise or hums that could change tone quality
Biological check
- can be completed using a device caked an bioacoustics simulator or with an individual with good stable hearing