Anesthesia Gas Machine Exam 1 Flashcards
Ideal Gas Machine
safe
efficient
economical
accurate
AANA Standards for CRNA practice includes (4)
- patient centered high quality consistent anesthesia care
- assist public to better understand role as CRNA’s
- foundation for CRNA practice
- modification of standards (always document modification and why)
which standard mentions equipment
standard 6
American Society for Testing and Materials (ASTM F1850-00)
- standard specifications for particular requirements for anesthesia workstations and their components
- published 2000, updated 2005
- workstations must have monitors that include: battery backup, high/medium/low priority alarms, required monitors, hanger yokes, PISS, electrical supply cord, flowmeters vaporizers
1L Liquid O2 = _____ gas O2
860L
pipeline supply, how much supply should be on hand at one given time
primary, secondary, minimum of one day reserve supply needed
standard flow rate for O2 and N2O through those systems
100L/min or greater (before regulator)
critical temperature
temperature below which a gas is going to be converted to liquid form by pressure
critical temperature of oxygen
-118c
what does drop in temperature do to molecules of a gas
slows movement (less KE)
pipeline tanks are:
temperature controlled, insulated, and pressurized
critical temperature of N2O
36.5c
pipeline supply specs (4)
- wall mounted
- connected to supply hoses
- filter and check valves
- check gauges
KISS
connects gas hose to wall outlet. unique key connected for each gas
DISS
diameter index safety system. each fitting is threaded and has different diameter and internal key unique for each gas. has valve that will shut off gas supply if its disconnected