Analytics Flashcards

1
Q

BGA - Why is it important to perform BGA analysis as soon as possible upon pulling the sample?

A

To avoid degassing and changes in sample temperature

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

BGA - What does a result reading “—-↑” or “—-↑↑” indicate?

A

The sample result exceeded the upper range limit for a selected parameter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

BGA - How often must the Wash/Waste Cartridge be replaced?

A

Every 10 days

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

BGA - When replacing both the Measurement and Wash/Waste Cartridge, which cartridge
should be replaced first?

A

Measurement Cartridge

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

BGA - What size syringe is used for BGA sampling analysis?

A

Up to a 30 mL syringe

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Osmometer - Why do we test just the supernatant for osmolality and metabolite analysis?

A

To avoid clogging samples lines of the analytical equipment with cell culture

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Osmometer - How often should you run the 290 reference standard test?

A

Once a day

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Vi-Cell Viability Analyzer - Why is it important to ensure that samples are thoroughly mixed prior to performing Cell
Density and Viability (CDV) analysis?

A

Cells settle over time; mixing ensures sample homogeneity in order to obtain an accurate cell count

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Vi-Cell Viability Analyzer - When must you perform manual sample dilution for CDV analysis?

A

When the total cell density goes above a certain limit (≥ 10 x 106 cells/mL)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Vi-Cell Viability Analyzer - How do you log in the samples into the Analyzer prior to testing the sample?

A

Click on the Log in Sample button

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Vi-Cell Viability Analyzer - When should the reagent pack be replaced?

A

When the reagent level meter under the diagnostics menu reads “Runs left: 0”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Turbidimeter - What should you be sure to do before inserting the sample into the device?

A

Gently invert the sample and apply silicone oil to the sample cell with the oil cloth

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Turbidimeter - What should you do if the 10 NTU fails to pass after a third attempt?

A

Calibrate the turbidimeter, as described in SOP000501

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Nova Bioprofile Flex - How long do new auto QC (AQC) Cartridges need to thaw?

A

At least 30 minutes prior to installation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Nova Bioprofile Flex - Why should the sample be spun down in a centrifuge?

A

To avoid clogging the device with cell culture. Only the supernatant will be used for sample testing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Nova Bioprofile Flex - How long should you shake the chemistry cartridge after adding the glutamine cups to the
reagent?

A

2 minutes to completely dissolve the glutamine powder