analytical techniques Flashcards
UV-VIS
detection and quantification of compounds
based on the ability of molecules to absorb light at wavelengths in UV and visible light region 180nm - 800nm
absorption of these wavelengths is characteristics of bonds and functional groups in molecules
wavelengths and energy
longer wavelength = lower energy
shorter wavelength = higher energy
UV VIS benefits
non destructive of compounds
simple and easy to operate
relatively low cost
used for both qualitative and quantitative analysis
extensive/complex sample prep not needed
used to test wide range compounds
example applications - chromatographic separations, quantification of protein/DNA, monitoring enzymatic reactions
principles of UV VIS
Absorption of light causes electron excitation from lower to higher energy orbitals.
Larger energy gaps between orbitals require more energy to excite electrons.
Greater energy jumps correspond to shorter wavelengths absorbed, while smaller gaps result in longer wavelengths absorbed.
UV VIS jumps between states
usually from pi to pi* (bonding pair of electrons pi bond to anti bonding pi*)
or from lone pairs electron (non bonding) = n to pi*
the energy difference between these jumps determines wavelength absorbed
after excitation the electron returns to ground state and releases the energy in from of light or heat
effect of conjugation - alternating pi and sigma bonds
Alternating pi and sigma bonds (electron delocalization) decrease the energy gap, requiring less energy for electron excitation, leading to absorption at longer wavelengths.
More pi bonds in a molecule make it easier for electron transitions from pi to pi* due to reduced energy required for the jump.
Delocalized electrons are more stabilized as they spread over atoms, reducing electron-electron repulsion.
homo and lumo electrons
homo = highest occupied orbital - contains highest energy electron
lumo - unoccupied energy level above homo electron
difference between these is called the band gap = transition energy. this is what determines wavelengths that are absorbed
more than one absorption peaks
Multiple Electron Transitions: Different types of transitions (e.g., pi to pi, sigma to sigma) require different amounts of energy, leading to multiple peaks.
Conjugation & Delocalization: Conjugated systems with multiple double bonds cause various transitions at different energy levels, creating several peaks.
beer Lambert law
intensity of light absorbed is quantified by beer Lambert law
transmittance = intensity of light out / intensity of light in
use this for measuring the concentration as it is difficult to weight out biological molecules accurately due to presence of water and salts and small quantities used
schematics of UV VIS
single beam instrument - monochromator determines wavelength
double beam instrument - uses reference and sample cell
UV VIS of proteins
chromophores are amide backbone with various side chains
dominant chromophore is amide group
- weak n->pi* transition at 220nm
- intense pi->pi* transition at 195nm
2nd absorption band around 280nm due to aromatic amino acids
pi to pi* peak
200-400nm
double bonds or benzene
e- from pi to pi*
as number of c=c increases, peak shift to longer wavelengths
n to pi* peak
300-600nm
molecules with lone pair e- adjacent to double bonds c=o n=o
energy required generally lower than pi - pi* so longer wavelength
peaks are less intense
sigma to sigma * peaks
below 200nm
molecules with single bonds only
requires high energy in UV region out of range
peak intensity
high intensity peak = high absorptivity
lower intensity = lower absorptivity
this relates to beer Lambert rule
broad peaks = overlapping transitions
narrow = more discrete specific electron transitions
uv vis of nucleic acids
chromophores
- nucleic bases
intense pi - pi* transitions
absorption spectrum of DNA
polynucleotides and nucleic acids absorb less per nucleotides than free ones
so depending on solvent, graph may look different
hypochromism
decrease in absorbance intensity
- due to molecular interactions and pi stacking
- structural changes
- electron delocalisation
example of hypochromism with heating and cooling
as heating, DNA becomes unstacked, interactions between bases decrease and absorbance increases
uv vis to monitor self assembly
by tracking changes in the absorption spectra as the molecules interact and form complexes or aggregates.
Absorption Changes: Self-assembly causes shifts or new peaks in absorption spectra as biomolecules interact.
Red/Blue Shifts: Assembly can cause red (longer wavelength) or blue (shorter wavelength) shifts in absorption peaks.
Intensity Changes: Peak intensity may decrease or increase due to changes in the molecular environment, like aggregation or complex formation.
fluorescent spectro
Fluorescence Emission: Measures light emitted after absorption by the sample.
Excitation: Sample is illuminated with light to excite electrons from the ground state to an excited state.
Excitation Wavelength: Based on the absorption characteristics of the fluorophore.
Vibrational Relaxation: After excitation, the molecule loses energy via vibrational relaxation before reaching a stable vibrational state.
Photon Emission: Once stable, the molecule returns to the ground state and emits a photon.
Stokes Shift: Emitted light has a longer wavelength than the excitation light.
Emission Detection: Detects emitted light and measures intensity across different wavelengths.
applications of fluorescence
detect presence of proteins
measure enzymatic activity
cell imaging
DNA concentration
biological events
advantages of fluorescence spectroscopy
high detected sensitivity - single molecule detection possible
high selectivity - 3 characteristic parameters, excitation wavelength, emission wavelength, lifetime
sensitive to molecular environment
sensitive to intermolecular interactions
both organic and inorganic species
fluorescence comes from …
arises from pi to pi* transitions - lots of aromatic rings and double bonds
- requires compounds with pi bonds and conjugated electrons
fluorescence quantum yield
number photons emitted/number photons absorbed
a higher quantum yield, a better compound for fluorescence - easier to detect
this is different from UV VIS as for that, interested in absorption of product
fluorescence spectrum - jablonski diagram
Excitation (Absorption): Molecule absorbs light, exciting electrons to vibrational levels of the first excited singlet state.
Vibrational Relaxation: In solution, the molecule rapidly loses vibrational energy to the solvent through collisions, converting energy into heat.
Fluorescence Emission: Molecule emits fluorescence from the lowest vibrational level of the singlet state.
Thermal Population: Higher vibrational levels can be thermally populated.
Transition to Ground State: The molecule then transitions from the excited singlet state to vibrational levels of the ground state, emitting light in the process.
stokes shift
Definition: The difference in wavelength between the excitation light (absorbed) and the emitted fluorescence (after energy loss).
Cause: After excitation, a molecule loses some energy through vibrational relaxation before emitting light, resulting in longer wavelength (lower energy) fluorescence compared to the excitation light.
Key Point: Stokes shift is observed because the emitted light has less energy than the absorbed light.
will lie more to the right of graph than absorption, often mirror images
emission spectrum - The emission spectrum tells you which wavelengths are emitted after excitation.
Fixed Excitation Wavelength: The sample is excited with a specific wavelength of light.
Emission Monochromator: A monochromator scans the emitted light at different wavelengths.
Fluorescence Intensity: The intensity of fluorescence is recorded as a function of the emission wavelength.
Spectrum Overview: The emission spectrum shows where fluorescence occurs and the peak intensity at the wavelength with the highest emission.
excitation spectrum - The excitation spectrum tells you which wavelengths can excite the fluorophore.
Fixed Emission Wavelength: The fluorescence emission is detected at a fixed wavelength.
Scan Excitation Monochromator: The excitation light is varied by scanning the excitation monochromator.
Fluorescence Intensity: Fluorescence intensity is recorded as a function of excitation wavelength.
Absorption Equivalent: The excitation spectrum is equivalent to the absorption spectrum of the fluorophore at the emission wavelength.
Information Provided: It gives information about the absorbance properties of the fluorophore.
intrinsic fluorescence of proteins
intrinsic fluorphores are the aromatic amino acids
- tryptophan
- tyrosine
- phenylalanine
collisional quenching
Definition:
Collisional quenching occurs when a molecule in an excited state loses energy due to a collision with another molecule, instead of emitting light (fluorescence or phosphorescence).
What happens? Excited molecule + quencher molecule → Energy lost as heat, no light emitted.
Cause? Physical collision between molecules.
Common quenchers: Oxygen, iodide, or other molecules that can absorb the energy.
Where? Happens in solutions or gases where molecules move freely and collide often.
use fluorescence to monitor protein changes
if protein is buried in a protein and less exposed to solvent - blue shift and peak/graph moves to the left
if protein becomes more exposed to solvent then red shift and peak/graph moves to the right
quenching
causes the glow to decrease
effect of calcium binding on calmodulin structure
ca2+ binding protein
activates proteins by binding, causing conformational changes - opens it up the protein
solvent effect
Definition:
how the polarity or properties of the solvent influence the fluorescence intensity, emission wavelength, and lifetime of a fluorophore (fluorescent molecule).
What changes?
Emission wavelength: Polar solvents stabilize the excited state, causing red-shifting (longer wavelengths).
Fluorescence intensity: Can increase or decrease depending on solvent interactions.
Lifetime: Solvent polarity and quenching affect how long the molecule remains fluorescent.
Why does this happen?
Polarity: A polar solvent stabilizes the fluorophore’s excited state differently than the ground state, altering energy levels.
Hydrogen bonding or dipole-dipole interactions: These interactions can either enhance or quench fluorescence.
Example:
A fluorophore like anthracene emits at different wavelengths in ethanol (polar) vs. hexane (non-polar).
foster resonance energy transfer FRET
energy is transferred non-radiatively (without emitting light) from an excited donor fluorophore to an acceptor fluorophore through dipole-dipole coupling.
What happens?
A donor fluorophore absorbs light and enters an excited state.
Instead of emitting light, the donor transfers energy to a nearby acceptor molecule.
The acceptor can then fluoresce or dissipate the energy in other ways.
Applications:
Measuring molecular distances (e.g., protein folding or interactions).
Biosensors to monitor cellular processes.
FRET
observe interactions in cells transferring energy without light emission
donor and acceptor molecule need to be close
see emission from acceptor
unpolarised light
polarisation potential
random oscillation- all directions
common sources
pass through polarising filter to become polarised
polarised linear light
single oscillation plane
through polarising filter
scattering
used in LCD screens and sunglasses
circular polarised light
direction of oscillation plane is circular, helical
rightly circular polarised = clockwise rotation
left = anticlockwise rotation