Advanced MR Imaging Flashcards
The HEIGHT of a peak on an MR spectrum correlates to the _________ of a substance that was detected
Amount
Which of the following is NOT a metabolite found in a 1H spectrum of normal brain tissue?
Succinylcholine
The final result in a spectroscopy study is:
A spectrum of specific detected metabolites
Contrast in Diffusion Weighted Imaging (DWI) depends on:
The degree to which molecular mobility is impaired (water restriction)
The sensitivity of the sequence to Brownian motion
The direction of water diffusion
Common problems in DWI include
Bulk patient motion
Low signal-to-noise ratio
Sensitivity to magnetic susceptibility when using echo planar readout
: In DWI sequences, ADC stands for:
Apparent Diffusion Co-efficient
The final result in a perfusion study is:
A set of calculated images which indicate various flow characteristics
: ____________ can be described as water molecules moving the same way in every direction.
Isotropic diffusion
____________ can be described as water molecules moving in a preferred direction
Anisotropic diffusion
The imaging technique that employs a very rapid pulse sequence while the patient is at rest or performing a task is
Functional MRI
Image 80 is an example of:

Single voxel spectroscopy
Letter A in Image 80 is pointing to the:

Amount of the substance/metabolite detected
Letter B in Image 80 is pointing to the

Frequency of the substance/metabolite detected
Image 81 is an example of:

Multi voxel spectroscopy
Letter A in Image 81 is pointing to the

Amount of the substance/metabolite detected
Image 81’s acquisition required the use of IV gadolinium contrast to obtain the image

False
Image 93 is an example of:

Diffusion tensor imaging
: Image 94 is an example of

Diffusion tensor imaging (tractography)
Image 95 is an example of:

Diffusion tensor imaging (tractography)
Image 96 is an example of:

Diffusion tensor imaging (tractography)
Image 97 is an example of

MR perfusion
Image 97 required the use of IV gadolinium contrast to obtain the image and data shown?

True
: Image 98 is an example of

MR perfusion
The echoes collected from a 256/256 Sagittal Cervical spine have a frequency range 14.1 kHz above and below the central frequency, 1.5Tesla system with 220 Hz chemical shift between fat and water, the vertebral body signal will shift _____ pixels relative to the signal in intervertebral disk
- 0
- 2 x 1000 = 28200 Hz; 28200 ÷ 256 = 110; 220 ÷ 110 = 2 pixels shift between fat and water
What would the pixel shift be in a sequence acquired at 1.5Tesla (220 Hz) with a 224 matrix and a BW of 26.8 kHz?
1.8
26.8(bandwidth in kHz) x 1000 to convert into Hz = 26800
26800 ÷ 224(matrix) = 120 Hz frequency difference per pixel
220(chemical shift) ÷ 120 = 1.8, the fat image will be shifted 1.8 pixels along the frequency/readout axis relative to water