99m-Tc RP QC P III Flashcards
1
Q
Solid-Phase extraction
A
- RCP test
- pre-packaged cartridges
- Disposible
- may be reused
- Use solvents of different polarities to separate different chemical species
2
Q
Sep-pak QC procedure
A
- Sample Preparation (0.1 ml of RP in tuberculin syringe)
- Condition Sep-Pak ® (Wet/Activate)
▪ 1-2 mL Ethanol, Methanol, HCL, Acetonitrile, Saline or Sterile Water
(solvent) - Add 0.1 ml of RP sample to the long end (inlet) of the Sep-Pak ®
- Rinse Sep-Pak ®
▪ Ethanol , Methanol, HCL, saline (depending on RP)
▪ Collect each in a separate falcon tube (Fractions) – Bound, Free, HR
▪ Push 10 ml of air through the Sep-Pak ® after final rinse - Assay components (Falcon tubes & Sep-Pak ® individually)
- Calculate % Radiochemical Purity according to Package Insert
3
Q
Considerations for sep-pak QC
A
- accuracy vaires with flow rate and sample load
- Flow rate: 15-20ml/min
- FR too high- no interaction
- Backsplash- contaminations and falsely low RCP
- RP loaded in wrong end
- wrong sep-pak used
4
Q
RCP limits for radiopharmaceuticals
A
- HMPAO: >/= 80%
- DMSA: >/=85%
- Sulfur colloid: >/= 92%
- others >/= 90%