two areas of activity designed to ensure best possible diagnosis at acceptable patient radiation dose and minimum cost
quality assurance (QA)
quality control (QC)
controlled by health care organizations / agencies
quality assurance (QA)
QA program involves:
monitor patient scheduling
reception
preparation
answering question/queries
image interpretation
QA activities are principally the responsibility of?
radiologist and imaging service management
more tangible and obvious than QA
Quality Control (QC)
designed to ensure that radiologist is provided with an optimal image produced through good equipment performance and resulting in minimal patient radiation dose
QC
ensure that health care organizations are committed to providing high-quality services and care
QA
begins with the x-ray imaging systems used to produce the image and continues with the routine evaluation of image-processing facilities
QC
concludes with a dedicated analysis of each image to identify deficiencies and artifacts (along with their causes) and to minimize re-examinations
QC
Each new piece of radiologic equipment, whether it is x-ray producing or image processing, should be acceptance tested before it is applied clinically
QC
QC program consists of 3 steps
acceptance testing
routine performance monitoring
maintenance
QC are principally the responsibility of?
medical physicist
general purpose radiographic units have a minimum total filtration of?
2.5 mm Al
xray field must coincide with the light field of the variable-aperture light-localizing collimator
collimation
these devices are automatic collimators that sense the size of the image receptor and adjust the the collimating shutters to that size
positive beam-limiting (PBL) collimators
spatial resolution of a radiographic imaging system is determined by?
focal-spot size
measured when new equipment or a replacement xray tube is installed
focal-spot size
3 tools used for focal-spot size measurement:
pinhole camera
star pattern
slit camera
difficult to use and requires excessive exposure time
pinhole camera
easy to use but has significant limitations for focal-spot sizes less than 0.3 mm
star pattern
standard for measurement of effective focal-spot size
slit camera
used for quality control of clinical X-ray machine
kilovolt peak calibration
measure exposure time on the basis of?
irradiation time of an ion chamber or photodiode assembly
assessed annually or more frequently if a component of the operating console or the high voltage generator has undergone major repairs
exposure timer accuracy
acceptable percentage of accuracy for exposure times of 10mc or less
20%
ability of a radiographic unit to produce a constant radiation output for various combinations of mA and exposure time
exposure linearity
determined by a precision radiation dosimeter that measures radiation intensity at various combinations of mA and exposure time
exposure linearity
Exposure linearity must be within ___ for adjacent mA stations
10%
ability of an exposure system to duplicate an exposure, time after time
exposure reproducibility