6.1 Instrumentation Flashcards
Which formula correctly describes the relationship between absorbance and percent transmittance?
A. A=2-log%T
B. A=logT
C. A=-log%T
D. A=2-%T
A
A=2-log%T
A Solution that has a transmittance of 1.0%T would have an absorbance of:
A. 1.0
B. 2.0
C. 1.0%
D. 99.0%
B
2.0
In absorption spectrophotometry:
A. Absorbance is directly proportional to transmittance
B. Percent transmittance is directly proportional to concentration
C. Percent transmittance is directly proportional to the light path
D. Absorbance is directly proportional to concentration
D
Absorbance is directly proportional to concentration
Which wavelength would be absorbed strongly by a red-colored solution?
A. 450 nm
B. 585 nm
C. 600 nm
D. 650 nm
A
450 nm
A green-colored solution would show highest transmittance at:
A. 475 nm
B. 525 nm
C. 585 nm
D. 620 nm
B
525 nm
Situation: A technologist is performing an enzyme assay at 340 nm using a visible-range spectrophotometer with a tungsten light source. After setting the wavelength and adjusting the readout to zero %T with the light path blocked, a cuvette with deionized water is inserted. With the light path fully open and the 100%T control at maximum, the instrument readout will not rise above 90%T. What is the most appropriate first course of action?
A. Replace the source lamp
B. Insert a wider cuvette into the light path
C. Measure the voltage across the lamp terminals
D. Change the wavelength to 335 nm
A
Replace the source lamp
Which type of monochromator produces the purest monochromatic light in the ultraviolet (UV) range?
A. A diffraction grating and a fixed exit slit
B. A sharp cut off filter and a variable exit slit
C. Interference filters and a variable exit slit
D. A prism and a variable exit slit
D
A prism and a variable exit slit
Which monochromator specification is required to measure the true absorbance of a compound having a natural absorption bandwidth of 30 nm?
A. 50-nm bandpass
B. 25-nm bandpass
C. 15-nm bandpass
D. 5-nm bandpass
D
5-nm bandpass
Which photodetector is most sensitive to low levels of light?
A. Barrier layer cell
B. Photodiode
C. Diode array
D. Photomultiplier
D
Photomultiplier
Which condition is a common cause of stray light?
A. Unstable source lamp voltage
B. Improper wavelength calibration
C. Dispersion from second-order spectra
D. Misaligned source lamp
C
Dispersion from second-order spectra
A linearity study is performed on a visible spectrophotometer at 650 nm and the following absorbance readings are obtained:
Concentration of Standard [absorbance]
10.0 mg/dL [0.20]
20.0 mg/dL [0.41]
30.0 mg/dL [0.62]
40.0 mg/dL [0.79]
50.0 mg/dL [0.92]
The study was repeated using freshly prepared standards and reagents, but the results were identical to those shown. What is the most likely cause of these results?
A. Wrong wavelength used
B. Insufficient chromophore concentration
C. Matrix interference
D. Stray light
D
Stray light
Which type of filter is best for measuring stray light?
A. Wratten
B. Didymium
C. Sharp cutoff
D. Neutral density
C
Sharp cutoff
Which of the following materials is best suited for verifying the wavelength calibration of a spectrophotometer?
A. Neutral density filters
B. Potassium dichromate solutions traceable to the National Bureau of Standards reference
C. Wratten filters
D. Holmium oxide glass
D
Holmium oxide glass
Why do many optical systems in chemistry analyzers utilize a reference light path?
A. To increase the sensitivity of the measurement
B. To minimize error caused by source lamp fluctuation
C. To obviate the need for wavelength adjustment
D. To reduce stray light effects
B
To minimize error caused by source lamp fluctuation
Which component is required in a spectrophotometer to produce a spectral absorbance curve?
A. Multiple monochromators
B. A reference optical beam
C. Photodiode array
D. Laser light source
C
Photodiode array
The half bandwidth of a monochromator is defined by:
A. The range of wavelengths padded at 50% maximum transmittance
B. One half the lowest wavelength of optical purity
C. The wavelength of peak transmittance
D. One half the wave length of peak absorbance
A
The range of wavelengths padded at 50% maximum transmittance
The reagent blank corrects for absorbance caused by:
A. The color of reagents
B. Sample turbidity
C. Bilirubin and hemolysis
D. The intrinsic absorbance of both the reagents and sample matrix
A
The color of reagents
A plasma sample is hemolyzed and turbid. What is required to perform a sample blank to correct the measurement for the intrinsic absorbance of the sample when performing a spectrophotometric assay?
A. Substitute deionized water for the sample
B. Dilute the sample 1:2 with a standard of known concentration
C. Substitute saline for the reagent
D. Use a larger volume of the sample
C
Substitute saline for the reagent
Which instrument requires a highly regulated direct current (DC) power supply?
A. A spectrophotometer with a barrier layer cell
B. A colorimeter with multilayer interference filters
C. A spectrophotometer with a photomultiplier tube
D. A sensitometer with a photodiode detector
C
A spectrophotometer with a photomultiplier tube
Which statement regarding reflectometry is true?
A. The relation between reflectance density and concentration is linear
B. Single-point calibration can be used to determine concentration
C. 100% reflectance is set with an opaque film called a white reference
D. The diode array is the photodetector of choice
C
100% reflectance is set with an opaque film called a white reference
Bichromatic measurement of absorbance can correct for interfering substances if:
A. The contribution of the interferent to absorbance is the same at both wavelengths
B. Both wavelengths pass through the sample simultaneously
C. The side band is a harmonic of the primary wavelength
D. The chromogen has the same absorbance at both wavelengths
A
The contribution of the interferent to absorbance is the same at both wavelengths
Which instrument requires a primary and secondary monochromator?
A. Spectrophotometer
B. Atomic absorption spectrophotometer
C. Fluorometer
D. Nephelometer
C
Fluorometer
Which of the following statements about fluorometry is accurate?
A. Fluorometry is less sensitive than spectrophotometry
B. Fluorometry is less specific than spectrophotometry
C. Unsaturated cyclic molecules are often fluorescent
D. Fluorescence is directly proportional to temperature
C
Unsaturated cyclic molecules are often fluorescent
Which of the following components is not needed in a chemiluminescent immunoassay analyzer?
A. Source lamp
B. Monochromator
C. Photodetector
D. Wash station
A
Source lamp
Which substance is used to generate the light signal in electrochemiluminescence?
A. Acridinium
B. Luminol
C. Dioxetane phosphate
D. Ruthenium
D. Ruthenium
Light scattering when the wavelength is greater than 10 times the particle diameter is described by:
A. Rayleigh’s law
B. The Beer-Lambert law
C. Mie’s law
D. The Rayleigh-Debye law
A. Rayleigh’s law
Which statement regarding nephelometry is true?
A. Nephelometry is less sensitive than absorption spectrophotometry
B. Nephelometry follows Beer’s law
C. The optical design is identical to a turbidimeter except that a helium-neon (HeNe) laser light source is used
D. The detector response is directly proportional to concentration
D. The detector response is directly proportional to concentration
The purpose of the nebulizer in an atomic absorption spectrophotometer that uses a flame is to:
A. Uniformly distribute the sample in the flame
B. Cause ejection of an outer shell electron
C. Reduce evaporation of the sample
D. Burn off organic impirities
A. Uniformly distribute the sample in the flame
A flameless atomic absorption spectrophotometer dehydrates and atomizes a sample using:
A. A graphite capillary furnace
B. An electron gun
C. A thermoelectric semiconductor
D. A thermospray platform
A. A graphite capillary furnace
When measuring lead (Pb) in whole blood using atomic absorption spectrophotometry, what reagent is required to obtain the needed sensitivity and precision?
A. Lanthanum
B. Lithium
C. Triton X-100
D. Chloride
C. Triton X-100
Interference in atomic absorption spectrophotometry caused by differences in viscosity is called:
A. Absorption interference
B. Matrix effect
C. Ionization interference
D. Quenching
B. Matrix effect
Which of the following is required when measuring magnesium by atomic absorption spectrophotometry?
A. A mercury vapor source lamp
B. A chopper to precent optical interference from magnesium emission
C. A neutral density filter
D. A 285-nm reference beam to correct for background absorption
B. A chopper to precent optical interference from magnesium emission