3. Central Venous Pressure and Cannulation Flashcards
Central Venous Pressure (CVP): The Waveform
Three upstrokes (the ‘a’, ‘c’ and ‘v’ waves) and
two descents (the ‘x’ and ‘y’) that relate to the cardiac cycle
‘a’ wave
‘a’ wave: this occurs at the end of diastole and is caused by increased atrial pressure
as the atrium contracts (occurs at end-diastole)
Atria
‘x’ (or ‘x”) descent:
‘x’ (or ‘x”) descent: this reflects the fall in atrial pressure as the atrium relaxes
relaX
‘c’ wave:
‘c’ wave: this supervenes before full atrial relaxation, and is caused by the
bulging of the closed tricuspid valve into the atrium at the start of isovolumetric right ventricular contraction.
(closed contraction)
‘x’ descent:
‘x’ descent:
this is a continuation of the ‘x” descent (interrupted by the ‘c’ wave) and
represents the pressure drop as the ventricle
and valve ‘screw’ downwards at the end of systole.
‘v’ wave:
‘v’ wave: this is the increase in right atrial pressure as it is filled by the venous return
against a closed tricuspid valve.
Venous Valve
‘y’ descent:
‘y’ descent: this reflects the drop in pressure as the right ventricle relaxes, the
tricuspid valve opens and the atrium empties into the ventricle.
‘y’ descent:
‘y’ descent: this reflects the drop in pressure as the right ventricle relaxes, the
tricuspid valve opens and the atrium empties into the ventricle.
Alterations
Any event that alters the normal relationship between these events will alter the
shape of the waveform.
For example, in atrial fibrillation the
‘a’ wave is lost;
in tricuspid incompetence,
a giant ‘v’ wave replaces the ‘c’ wave, the ‘x’ descent and the ‘v’ wave.
‘Cannon’ waves are seen when there is atrial contraction
against a closed tricuspid valve (as occurs at a regular interval if there is a
junctional rhythm, or at an irregular interval if there is complete atrioventricular
conduction block).
Indications:
CVC catheters are used for the monitoring of central venous pressures,
for the insertion of pulmonary artery catheters
and to provide access for haemofiltration
transvenous cardiac pacing.
Central venous lines also allow the administration of
drugs that cannot be given peripherally,
such as inotropes and cytotoxic agents,
and the infusion of total parenteral nutrition.
It is suggested that they can be used to aspirate air from the right side of the heart after massive air embolism, although very few anaesthetists have ever used them for this
purpose
Function of CVP monitoring – Volume
Function of CVP monitoring – intravascular volume:
the CVP is the hydrostatic pressure generated by the blood within the right atrium (RA) or the great veins of the thorax.
It provides an indication of volaemic status because the capacitance system,
including all the large veins of the thorax,
abdomen and proximal extremities,
forms a large compliant reservoir for around two-thirds of the total blood volume.
Hypovolaemia may be actual or effective, caused, for example,
by subarachnoid block or sepsis,
in which loss of venoconstrictor tone or venodilatation decreases venous return and reduces CVP.
A single reading
may be unhelpful, whereas trends are more useful, particularly when combined
with fluid challenges.
Function of CVP monitoring – right ventricular function
Function of CVP monitoring – right ventricular function: CVP measurements
also provide an indication of right ventricular (RV) function.
Any impairment of RV function will be reflected by the higher filling pressures that are needed to maintain the same stroke volume (SV).
Normal values
Normal values:
the normal range is 0–8 mmHg, measured at the level of the tricuspid valve.
The tip of the catheter should lie just above the RA in the superior vena cava.
CVP decreases:
CVP decreases:
if the blood volume is unchanged,
then the CVP will alter with changes in cardiac output (CO).
It will fall as the CO rises because the rate at which
blood is removed from the venous reservoir also increases
CVP increases:
CVP increases:
potential causes for an increase in CVP include a fall in CO
(the converse of the effect described previously).
Ventilatory modes may also cause the increase which is seen with IPPV, PEEP and CPAP.
The CVP also rises in response to volume overload,
if there is RV failure,
pulmonary embolus,
cardiac tamponade or
tension pneumothorax.
Rarer causes include obstruction of the superior vena cava
(assuming that the catheter tip lies proximally) and portal hypertension leading to
inferior vena caval backpressure.