2.1 Flashcards
5 Techniques
SEM TEM STM AFM XPS
SEM
Scanning electron microscopy
Refraction technique
Surface of specimen
Low energy 2ndry electrons emitted from surface of sample
- electron beam in vaccuum concentrated on small area and then rastered across the surface
- electrons detected and reconstructed into an image
morphology of crystals and biological shit not coloured
Resolution 1nm
TEM
Transmission electron microscopy
can see atoms pretty mch ish
transmission technique
Sample must be in vassum (e’s are deflected in the air)
sample can be affected by high energy electron beam
->damage to sample
HRES TEM - 0.5 A 4 resolution
STM
A conducting tip is brought close to a metallic/semiconducting surface.
Apply small bias electrons tunnel between then
Tunnelling current is proportional to local DOS at fermi level of sample
Variations in current as probe rasters the surface allows an image to be generated
STM general
Scanning Tunnelling microscopy v. high vacuum lateral res 1A depth 0.1 A produces weird looking 3d images etc tend to look glowy sample must conduct
AFM
Atomic force microscopy
insulating samples
resolution poorer than STM
a tip mounted on a cantilever rasters a surface
when tip is close to surface
forces between tip and surface result in a deflection of the cantilever
constant height scan or with feedback loop
can do in media
picture is grey white show atoms grey not defined
XPS
Xray photoelectron spectroscopy (XPS)
Bombard sample with x rays - many eletron processes can occur
one process is the ionisation of an electron in inner shell
detect the electrons emitted and calculate binding energy and other properties
E=hv-Eo
Eos can be calculated
There are characteristic range of Eo values from different shells for a given element
useful for surfaces but not bulk
as electrons detected are only from top 2-5 nm of surface