1990 MV Code of Practice Flashcards
Equation for NPL absorbed dose calibration service
D = RNd
Standard ambient conditions
T = 20 degrees celcius P = 760 mmHg/1013.25 mbar H = 50%
Depth of chamber used in calibration service & intercomparison
4-10 MV = 5 cm
12-19 MV = 7 cm
Field size used in calibration service & intercomparison
10x10
Water phantom size
Full scatter phantom, extending at least 5 cm outside the beam edges
Quality index for a MV beam
Tissue phantom ratio 20/10: the ratio of the corrected instrument readings with the chamber at 20 cm and 10 cm depths in a water phantom with a 10x10 field and constant source-chamber distance.
Why do we rarely correct for relative humidity?
Only 0.1% difference when relative humidity is at 10%-90%
How do you determine ion recombination for pulsed radiation?
Half-voltage technique. If chamber reading decreases by x% when the voltage is half its usual value, then the correction to be applied to the chamber reading is x%.
For NE 2561 chambers, if you don’t want to measure it then you can use, k = 1.0014 + 0.23p where p is the dose per pulse in the chamber in cGy.
How do you determine ion recombination for continuous radiation?
Ion recombination losses are often small here. For NE 2561 chambers, just use standard value of 1.0014.
What phantom can you use for intercomparison and what set-up method?
Water OR perspex phantom, sequential OR side-by-side
Merging all this information, what are the steps to performing an intercomparison between a field instrument and the SS?
- Make sure you have/measure the beam quality using TPR 20/10 to identify Nd for the SS
- Make sure you have/measure the ion recombination factor.
- Set up the two chambers sequentially or side-by-side, in a water or perspex phantom that extends at least 5 cm beyond the beam edges and has a field size of 10x10, with the chambers at the appropriate reference depth for the energy.
- Allow equipment to stabilise and phantom to come to room temperature. Take T and P.
- Obtain measurements, correct for FTP and ion recombination.
- Using D = RNd(SS) = RNd(F) to deduce the calibration factor for the field instrument.
What is the primary standard for MV dosimetry?
Graphite calorimeter