12-lead Acquisition Flashcards
1
Q
What is tue correct frequency for acquiring a 12-lead?
A
0.05-40Hz
2
Q
How do you find the accurate placement site of V1?
A
- Locate supra-sternal notch
- Run a finger down the manubrium until the ridge of meets the body of the sternum (Angle of Louis)
- The 2nd rib lies immediately adjacent to the Angle of Louis), find the intercostal space
- Move down two more intercostal spaces
- Place electrode immediately to R) of sternum
3
Q
Order the placement of chest leads
A
1st - V1, 4th intercostal space, R) of sternum
2nd - V2, “” L) of sternum
3rd - V4, 5th intercostal space mid-clav line
4th - V6, 5th intercostal space, mid-axillary
5th - V3, Midway V2-V4
6th - V5, Midway V4-V6