Concerning cyrogenic liquid systems, what volume of gas does 1 volume of liquid oxygen give you?

A

842 times its volume.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is the pressure inside a VIE?

A

10.5 bar

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Does oxygen require heating at the outlet of a cryogenic liquid system?

A

heating of gaseous oxygen is required at the outlet using a heat exchanger

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What is the length of onsite storage capacity requirement of VIE?

A

14 days.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Is nitrous oxide liquid in cylinders?

A

yes, up to 95% liquid

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Regarding pipes:
What are they made of?
What is the pipeline pressure of oxygen, N2O, medical air and surgical air?

A

Made of copper,

4 bar for all except surgical air which is 7 bar.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Difference between where Schrader sockets and NIST are used?

A
  1. Schrader sockets and probes are used to connect to terminal outlets for medical gases.
  2. NIST is used to connect to anaesthetic machines.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What flow does the emergency oxygen flush have to provide?

A

It should provide at least 35 L/min to match inspiratory flow rate

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What flow should suction be able to provide?

A

25 L/min is a required displacement capacity of suction system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

How do you check the performance of the suction system?

A

should be able to generate negative pressure of more than 500 mmHg in 10 seconds when tube is kinked.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Suction tubing should be off high/ low resistance and compliance?

A

low resistance and compliance.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

In a TEC 6 Vapouriser:
What is the purpose of the splitting ratio?
What is Desflurane heated to?

A

Splitting ration ensures saturation of fresh gas flow with anaesthetic vapour.

TEC 6 is heated to 39C.

23.5C is boiling temperature of Desflurane

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

How much CO2 will 1kg of soda lime absorb?

A

250 L

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What is the largest component of soda lime?

A

The largest component calcium hydroxide (80%)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What size is the scavenger connector?

A

30 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What information is required to calculate the remaining amount of N2O in a cyclinder?

A

MW of N2O (44)
Avogadro’s number (22.4)
cylinder’s weight
cylinder’s weight when empty

17
Q

In a TEC 6 vaporiser, what is the vaporising chamber pressurised to?

A

The vaporisation chamber is pressurised to approximately 200 kPa

18
Q

What type of vapouriser is a TEC 6?

How is it different from other vapourisers?

A

It is called measure flow vaporiser

It is different from ordinary TEC vaporizers to provide very accurate electronically controlled gas injection of Desflurane into carrier gas.

Incorporates a differential pressure transducer

19
Q

What is the pressure across APL valve limited to?

A

60cm H20

20
Q

What size of connectors are generally used in a breathing system?

A

15 mm and 22 mm,

except for scavenging which uses 30 mm.

21
Q

What does the oxygen failure alarm depend upon?

Does it depend upon an electrical source for the alarm?

A

Activation is dependent solely on oxygen supply pressure. It is coupled to a valve that interrupts the flow of anaesthetic gases

Electrical source not required for the alarm.