Module 6 Final Flashcards
____ took the first intraoral radiograph.
Walkhoff
_____ was a professor of physics, and discovered X-rays.
Roentgen
The open-ended tube, commonly referred to as the cone is called the:
Position Indicator Device (PID)
The ____ is a negative electrode that produces electrons.
Cathode
The sharpness or definition of the radiograph is determined by the:
Focal Spot
Radiation trauma that causes changes to the ____ cells will affect future generations in anomalies or birth defects.
Genetic
The dose to which the body tissues are exposed, measured in terms of its estimated biological effects, is the:
Sv.
the portion of the primary beam that is limited by the collimator inside the position indicating device is known as the:
Useful Beam
The aluminum material that is placed in the path of the beam of radiation to remive many soft rays or longer wavelengths that are undesirable is the:
Filter
The recommended postion for the operator during x-radiation exposure is:
Behind a leaded door at a distance of at least 6 feet from the central beam.
____ is the most common setting for the milliamperage.
10-15 mA
____ is the setting on the control panel that regulates the force with which the electrons move across the tube gap.
Kilovoltage (kVp)
if the operator increases the exposure time, the film will appear:
Darker.
There are several tyes of film holders available, including Styrofoam, Stabe bite-block, ____, EEZEE grip, Uni- BIte, and EndoRay.
XCP
The type of film placed directly into the mouth for exposure is called a/n ____ film.
Intraoral
The two most common extraoral films used to investigate the jaws and skull are the cephalometric and the ____ film.
Panoramic
The fastest speed film is:
F-speed.
The safe light within the dark room must be at least ____ feet away from the working surface during film processing.
Four
The floaating-type thermometer is placed in the ____, and should be periodocally checked for radiographic processing.
Developer
The chemicals used to process films should be changed approximately every ____ weeks.
3-4
The processing tanks must have a/n ____ to prevent the film from light exposure, oxidation, and evaporation of chemicals.
Metal lid
When processing film at higher temperatures, the developing time is:
Decreased.
Dark films are the result of:
overdeveloping and overexposing.
Too short a fixing time can result in:
Partial loss of the image.
When the developing solution becomes exhausted, processed films appear:
Light.
All dental x-ray equipment made after 1974 is required by federal regulations to meet safety standards for minimum filtration and x-ray equipment control settings. The three settings required for inspection are ____, mA, and time.
kVp
Federal legislation requires dental auxilliaries who will be exposed to radiography to:
Successfully pass a radiology course as required by the individual’s State Board of Dentistry.
Exposed radiographs:
Are opened without touching hte contaminated dental film with ungloved hands.
Film holding devices and alignment devices are classified as ____ instruments, and must be sterilized.
Semi-critical
The American Academy of Oral and Maxillofacial Radiology recommends the ____ technique for intraoral dental radiographs.
Parallelling