Definitions and Formulas Flashcards
Table Movement / Beam Width = ?
Formula for Pitch
Commonly used to describe table movement.
Slice Number x Thickness = ?
Formula for Beam Width
16 slice scanner x 2.5mm thickness =
40 beam width
CDTIw(eighted) / Pitch = ?
CDTI vol(ume metric)
Radiation absorbed by patient’s body.
Expressed in Gy.
CT Dose Index
(CTDI)
An approximate measure of the radiation dose received in a single CT section or slice.
With a 16 slice MDCT at 0.625mm thickness and a table feed of 15mm per rotation, what is the pitch?
15mm / (16 x 0.625mm)
1.5 pitch
A 1% increase in attenuation = ———— ————- in HU.
10 increase
The pixel value, given in HU, at the center of the WW.
Window Level (WL)
Controls the brightness (density) of the CT image.
The range of pixel values assigned a shade of gray in the displayed CT image.
Window Width (WW)
Controls the contrast of the CT image.
What is the distance covered for a MDCT helical scan sequence with a 30 second total acquisition time, a 1 second rotation time, 5mm slice thickness, 16 slices per rotation, and a pitch of 1?
Pitch * Acquisition Time * 1/Rotation Time * Beam Width = Scan Coverage
1 * 30 * 1/1 * (5 * 16) = 2400mm
Formula for calculation of pixel size
FOV / Matrix size = pixel size
matrix is usually 512