Breathing systems Flashcards
What should always be done to the expiratory/APL valve before attaching a patient?
OPEN
Closed = high pressure, lung damage!
When doing a leak test, what should be done to expiratory/APL valve?
Closed
RE-OPEN BEFORE ATTACHING PATIENT
Re-breathing circuits can be used in patients over what size?
25kg
What is the flow rate for rebreathing circuits?
10-50mls/kg/min
(Or
2.5-3L/min until anaesthetic stable
Then 0.5-1L/min)
What is the flow rate for non-rebreathing circuits?
200ml/kg/min
Consider circuit factor
What is the circuit factor for Mapleson A (Magill)? What weight must be the animal be?
Circuit factor 1
Weight >10kg
What is the circuit factor for Mapleson D? What weight must be the animal be?
Circuit factor 2-2.5
Weight <10kg
What is the circuit factor for Mapleson E (Ayre’s T-piece)? What weight must be the animal be?
Circuit factor 2.5-3
Weight <10kg
What is the circuit factor for Mapleson F (Jackson-Rees)? What weight must be the animal be?
Circuit factor 2.5-3
Weight <10kg
What is the circuit factor for parallel lack? What weight must be the animal be?
Circuit factor 1-1.5
Weight >10kg
What is the circuit factor for Coaxial Bain (modification of Mapleson D)? What weight must be the animal be?
Circuit factor 2-2.5
Weight >10kg